Ensure that the Tubing Guide Arm is not missing and that it opens when the Pump module
door opens. A Tubing Guide Arm failure can result in a nuisance air-in-line alarm. If there is a
failure, send the Pump module to CareFusion for repair
NOTE:
There are no changes to set loading, door closure, or latching.
BD Alaris System User Manual Addendum
.
Enhancement of
upper tubing
fitment
Enhancement of
upper tubing fitment
to support set
loading
Load First text
Reinforces the
placement of the
administration set
into the upper
fitment as the first
step of set
loading.
Air-in-Line
Detector
Tubing Guide Arm
This feature secures
the administration set
in place within the Air-
in-line detector