Appendix B: Tips on How to Calibrate the Keysight N9912A FieldFox
RF Analyzer
S
Calibration
11
1. Set up the connections as shown in Figure 12, and then connect the calibration kit to the SMA
connector. The suggested calibration kit for the following calibration procedure is the Keysight
85033E.
2. On the N9912A front panel, press Mode. Use the softkey to select NA, and then select S
3. Press Cal (button #5) on the N9912A front panel. Use the softkeys to select [O,S,L] for the
Calibration Type.
4. Select Start Cal and press the corresponding softkey Open –F–.
5. Connect the OPEN CIRCUIT calibration kit to the SMA connector and press Measure.
6. Upon completing the measure, press the corresponding softkey Short –F–.
7. Connect the SHORT CIRCUIT calibration kit to the SMA connector and press Measure.
8. Upon completing the measure, press the corresponding softkey Load –F–.
9. Connect the TERMINATION calibration kit to the SMA connector and press Measure.
10. Upon completing the measurement, press Finish to complete the O,S,L calibration. The
calibration can be repeated again at any time in the event that the calibration kit was not properly
connected during the calibration process.
ME1310 Antenna and
Propagation (3D)
Figure 12 – Setup for S
http://dreamcatcher.asia/cw
calibration
11
.
11
Lab 1 - 15/18