Ambulatory Blood Pressure Monitor TM2441-EX,
Instruction Manual, Corrections and Additions
Thank you for purchasing the Ambulatory Blood Pressure Monitor TM2441-EX. The Instruction Manual that
accompanies the product has been corrected and had content added. Please read it in conjunction with the
details shown below.
Instruction Manual Page 27 6.1. Automatic Blood Pressure Measurement (A-BPM)
The following additions have been made.
6.1.4. Saving Environmental Data
When the AUTO switch is "ON", environmental data are only saved during automatic
measurement operations.
Environmental data are saved for up to 12 days, but in this case, it will take about
50 minutes to import the data into the ABPM Data Manager.
Instruction manual Page 28 6.1. Automatic Blood Pressure Measurement (A-BPM)
The following changes have been made.
An initial pressurization value can be set beforehand.
160, 180, 210, 240, 270, AUTO [mmHg]
If AUTO is specified to the initial pressurization value, the
pressurization value is selected automatically. Factory
setting is 180 mmHg. Refer to "8.2.3. Initial
Pressurization Value" for information on how to select an
initial pressurization value.
Wrong
Addition
Caution
The initial pressurization value for automatic blood pressure
measurement is AUTO.
In the case of AUTO, the appropriate pressurization value
is set automatically.
1
Correct
1WMPD4004496