Function / Meaning

Symbols

Turns the device on and off.
Guide to install battery
Serial number
SN
Date of manufacture
Type BF: Device, cuff and tubing
are designed to provide special
protection against electrical
shocks.
Displays when measurement is in
progress and flashes when the
pulse is detected.
Irregular Heartbeat indicator. (I.H.B.)
The indicator that detects irregular
heartbeats or any excessive body
movement during the measurement.
Flashes when there is air
remaining in the cuff at power up.
Flashes at the completion of the
measurement until exhaust is
complete.
Flashes when the pressure is not
enough for the measurement.
The battery power indicator during
Full Battery
measurement.
The battery is low when it blinks.
low Battery
Unstable blood pressure due to
movement during the measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during inflation.
err CUf
The cuff is not fastened correctly.
pUl err
The pulse is not detected correctly.
The EC directive medical device
label
Symbols
ENGLISH 4
Recommended Action
Measurement is in
progress -- remain still.
Press the exhaust valve
button to exhaust air from
the cuff.
Pump air into the cuff with
the rubber bulb.
Replace the battery with a
new one, when the indicator
blinks.
Try measuring again.
Remain very still during the
measurement.
Fasten the cuff correctly,
and try the measurement
again.