AND UB-542 Руководство по эксплуатации - Страница 18
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Монитор артериального давления AND UB-542. AND UB-542 19 страниц. Wrist digital blood pressure monitor
Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you
cannot fix the problem using the troubleshooting instructions, contact the authorized dealer in your area or our
customer service department. The A&D customer service will provide technical information, spare parts and units
to authorized dealers.
The device was designed and manufactured for a long service life. However it is generally recommended to have
the device inspected every 2 years, to ensure proper functioning and accuracy. Please contact the authorized
dealer in your area or A&D for maintenance.
Type
UB-542
Measurement
Oscillometric measurement
method
Measurement
Pressure: 0 – 299 mmHg
range
Pulse:
40 – 180 beats / minute
Measurement
Pressure: ±3 mmHg
accuracy
Pulse:
±5%
Power supply 2 x 1.5V alkaline batteries (LR03 or AAA)
Approx. 250 measurements, when AAA Alkaline
Number of
batteries are used, with pressure value of
measurements
170 mmHg at room temperature of 23°C
Wrist
13.5 – 21.5 cm
circumference
Internally powered ME equipment
Classification
Continuous operation mode
Note: Specifications are subject to change for improvement without prior notice.
EMC table information is listed on our website: http://www.aandd.jp/products/manual/medical/emc_en.pdf
Maintenance
Technical Data
Applied part Cuff Type BF
Useful life
Clinical test
EMC
Memory
Operating
condition
Transport/stor
age conditions
Dimensions
Weight
English 17
Device: 5 years (when used six times a day)
According to ANSI / AAMI SP-10 1992
IEC 60601-1-2: 2007
Last 90 measurements
+10°C to +40°C / 15%RH to 85 %RH
800hPa to 1060hPa
-20°C to +60°C / 10%RH to 95 %RH
Approx. 56 [W] x 88 [H] x 18 [D] mm
Approx. 103 g, excluding the batteries