Andersen Sterilizers Anprolene AN75i Мануалы по установке

Просмотреть онлайн или скачать pdf Мануалы по установке для Лабораторное оборудование Andersen Sterilizers Anprolene AN75i. Andersen Sterilizers Anprolene AN75i 2 страницы. Installation options

Andersen Sterilizers Anprolene AN75i Мануалы по установке
Recommended Installation
When possible, the Anprolene Sterilizer
should be located near an outside wall,
and exhaust directly to the outside by
means of ¾" (1.9cm) I.D. tubing.
Through wall mounting kits are available
from Andersen. Note: Sterilizer must be
placed in an area that can maintain 68°F
or above room temperature.
Alternate Option 1
Through the roof applications are best accomplished by installing 2" SCH
40 PVC using a roof vent boot and stabilizing the PVC to extend just below
the ceiling. The ¾" I.D. to 2" adapter cap is available from Andersen.
Note: 2" SCH 40 PVC exhaust can run up to 800' depending on path -
bends/90° turns will reduce distance.
Ceiling
Quick
release
connector
Roof installation with vent boot
2" PVC pipe
Adapter (1)
Hose
Installation Guidelines
Anprolene sterilizer installation options
Quick
release
connector
Exterior*
2" PVC pipes
Exterior*
Roof
Quick
release
connector