2.3 TRANSMITTER ELECTRONIC "ZERO" CALIBRATION
2.3 TRANSMITTER ELECTRONIC "ZERO" CALIBRATION
2.3 TRANSMITTER ELECTRONIC "ZERO" CALIBRATION
2.3 TRANSMITTER ELECTRONIC "ZERO" CALIBRATION
2.3 TRANSMITTER ELECTRONIC "ZERO" CALIBRATION
Upon installation of a new Factory Calibrated unit, and as part of routine maintenance, a
Zero calibration adjustment must be performed. The transmitter ZERO, (signal output with
no pressure applied to the diaphragm), is 4.00 mA. Although the calibration may be
performed without additional tools, testpoints have been provided for monitoring the mA
output signal from the transmitter. For maximum accuracy we recommend performing a
sensor zero about three (3) weeks after initial installation, or following several heat/cool
cleaning cycles.
See Figure 2-4, Zero Calibration Procedure, for the location of the Zero switch. Depressing
for 5-8 seconds automatically "zeros" the output.
CAUTION:
• Field wiring MUST be complete – loop power (12-40 VDC) applied
• Transmitter MUST be installed in vessel
• Verify NO product contact to diaphragm
• Vessel MUST be vented to atmosphere
• DO NOT depress SPAN switch
• MODE switch set to RUN MODE
Once Zero calibration has been performed, sensor output will return to 4.00 mA. This is the
proper output to signify an empty vessel – ZERO calibration is complete.
NOTE:
No adjustment to the SPAN is necessary. ZERO and SPAN settings are non-
interactive, having no effect on each other.
FIGURE 2-4 ZERO CALIBRATION PROCEDURE
FIGURE 2-4 ZERO CALIBRATION PROCEDURE
FIGURE 2-4 ZERO CALIBRATION PROCEDURE
FIGURE 2-4 ZERO CALIBRATION PROCEDURE
FIGURE 2-4 ZERO CALIBRATION PROCEDURE
NOTE: RECEIVER AND METER CONNECTIONS
SHOWN WITH
CAP REMOVED
OMITTED FOR CLARITY
REFER TO FIGURE 2-3 FOR HOOKUP
DEPRESS
ZERO/-
SWITCH
TO INITIATE ZERO CALIBRATION
HOLD FOR 5-8 SECONDS TO
COMPLETE ZERO CALIBRATION
MODE SWITCH
SET TO RUN MODE