Section 7 - Troubleshooting

HM FLOWMETER TROUBLESHOOTING GUIDE
PROBLEM
CAUSE
No output or
erratic output
HT75000
Poor
accuracy/
repeatability
Failed Probe (no output)
Probe not seated
Moisture and/or corrosion
in probe-to extension
cable connector
Flow rate too low
Obstructed Rotor
Incorrect wiring
Bad cable splice
No scale factor entered
Signal interface problem
Incorrect coil resistence
Air in system
Process flow rate out of
meter flow range
Obstructed Rotor
Probe not seated
Poor meter installation
Poor system hydraulics
Incorrect scale factor
Electrical noise
Changing product
characteristics density
and/or viscosity
Inaccurate proving
standard
Operating error – poor
testing method
Rotor wear
Damaged housing
ACTION
Replace Probe
Check probe: must be threaded to bottom
Disconnect extension cable at probe for
visual check, dry or clean if required
Check meter specification for flow range
Check rotor (should spin freely)
Check manual for correct wiring configuration
Check color code and continuity integrity
Check readout device scale factor
Check signal input specs of receiving device
Check black/white signal wires for 975Ω
Add air elimination device
Check meter specs for flow range info (adjust
flow rate or change-out meter)
Check rotor (should spin freely)
Check probe: must be threaded to bottom
Locate meter in low spot (see installation)
Backflow? (Install check valve) Valve closing
time?
Check: scale factor, K factor and unit of
measure
Check: shielding, separate signal from AC
wires
Check: product weight, viscosity, scale
accuracy, control temp. Multiple products?
Verify accuracy of proving standard (vessel/
scale)
Observe test procedure
Remove and inspect (figure: 6.4)
Remove and inspect (figure: 6.5)