Atlanta BYK 7300 Глоссарий - Страница 10

Просмотреть онлайн или скачать pdf Глоссарий для Измерительные приборы Atlanta BYK 7300. Atlanta BYK 7300 20 страниц. Explanation of terminology

Atlanta BYK 7300 Глоссарий
spectro2profiler
BYK-Gardner GmbH
Plateau parameter
... controls the truncation of the crests of the hills. This data manipulation af-
fects the watershed segmentation on the hills. The data for the 3D analysis is
not affected. The parameter range is from 0.0 to 1.0. The default setting is 1,
which is also used in the instrument firmware.
If plateau parameter = 1, no truncation will be done. If plateau parameter = 0,
the crests of the hills will be flattened down to their mean height. The images
underneath show the truncated topography as red line and the segmentation
borders as green lines, illustrated in abstracted form.
Examples:
The following images show the truncated topography as red line and the seg-
mentation borders as green lines, illustrated in abstracted form.
Plateau = 1.0
Plateau = 0.3
Over Segmentation
Good Segmentation
Hill size threshold
... defines the height of the hill intersection for the hill size calculation. The
range is from -1.0 to +1.0. The default setting is 0, which is also used in
the instrument firmware.
Hill size threshold = 0 means that the mean height of the topography is used as
threshold. A value of -1.0 means that the lowest point is used as threshold. This
selects all data as hills. A value of +1.0 generates the opposite output. The rel-
ative threshold parameter ensures the comparability between different
samples.
Examples:
Hill sections are indicated as red areas. The green lines are the watershed seg-
mentation result. The abstracted illustrations above show the intersection
marked as red line and the segmentation borders as green lines.
10
283 025 846 - 2008