If the pipette malfunction persists after carrying out the
above steps, send the pipette to your A.B.T. service
representative.
Before returning the pipette, make sure that it is free from
any chemical, radioactive or microbiological contamina-
tion that might pose any danger during transportation and
repairing. Clean the pipette as thoroughly as possible.

10 - CLEANING AND STERILIZATION

Cleaning
External surfaces of the pipetting pushbutton, the ejector
pushbutton, the handgrip, the shaft nut and the adjust-
ment knob may be cleaned using a cloth dampened in
isopropyl alcohol. The remaining parts removed from the
pipette during pipette disassembly may be washed with
distilled water or isopropyl alcohol.
Sterilization:
The pipette can be sterilized in the autoclave at 121°C for
20 minutes. After sterilization, the pipette should be dried
and cooled to room temperature.
It is recommended:
- to sterilize the pipettes in autoclave with an initial vacu-
um and drying cycle,
- prior to sterilization unscrew the shaft nut slightly in the
AB10-AB1000 pipettes. After autoclaving these parts
should be screwed tight again.
The precision of the results should not alter if the pipetting
process and autoclaving are carried out as descri bed
in this manual. Because a slight change in the accuracy
of the dosage may occur, it is recommended to:
- check the calibration of the pipette after the initial first,
third and fifth autoclaving cycles and then after every 10
autoclaving cycles.

11 - PIPETTE KIT

The pipettes are delivered in the kits including:
• Pipette
• Instruction manual
• Calibration key
• Lubricant
• Ejector cap (10000 μl pipette)
13

12 - SPARE PARTS

All the spare parts and accessories indicated in Fig. 1, 3, 4:
A: Pipetting pushbutton
B: Adjustment knob
C: Shaft
D: Ejector
F: Shaft nut
G: Piston assembly
I: O-ring
J: Seal
K: Calibration key
N: Ejector pushbutton
Spare parts and accessories can be ordered from your
A.B.T. representative (type of pipette and name of the
part for this pipette should be specified).
Warning: The replacement of the piston requires
conducting of calibration procedure according to
section 8.
.
All rights reserved. Product described in this manual is subject
to availability and technical modification. Errors excepted.
A.B.T. reserves the right to improve, enhance or otherwise modify
its products without prior notification.

ENGLISH

A1: Pushbutton
A2: Knob
14