Ankersmid Sampling AHL Series Руководство пользователя - Страница 8
Просмотреть онлайн или скачать pdf Руководство пользователя для Нагреватель Ankersmid Sampling AHL Series. Ankersmid Sampling AHL Series 18 страниц. Heated sample lines
5. If the heated hose is laid outdoors is must be protected against wind, because it would
otherwise be subjected to cooling and may then not be able to attain the desired
temperature. Likewise, the heated hose should not be exposed to rain or direct solar
radiation for any length of time. In such cases suitable covering must be provided for
protection.
6. Observe all the following installation guidelines (wrong-right chart). These guidelines
were drawn up on the basis of many years of experience and fault analysis to avoid the
most common causes of failure.
7. Do not use the fitting to pull the heated hose. The fittings are resistant to pressure, but
more susceptible to tension forces.
Neither should the connecting cable be used for pulling the heated hose because it is not
suitable for this.
8. Are there any special rules or regulations applicable at the place of installation and has
the installation taken this into account?
9. The operator/customer must check if the material that comes into contact with the
media is resistant to the media to be heated.
10. Always monitor the first heating-up phases of the heated hose very carefully so that
any possible faults can be detected and, where appropriate, remedied at an early stage.
11. Check whether any nearby objects, plant components or other parts could cause
damage to the heated hose or impair its function and, where applicable, remove or
rearrange such items.
12. Any accessible (exposed) electro-conductive parts must be included into the
equipotential bonding measures.
The standard lines, temperature controlled by ATC-temperature controller, are equipped
with round connector plugs:
7-pin plug (max. 10A)
Colour code:
Temperature sensor (standard Pt100) = yellow
Power supply (standard 230VAC)
Ankersmid Sampling
Neerlandweg 21
2610 Wilrijk
Belgium
5-pin plug (>10A, max. 20A)
= brown (L) / blue (N)
User Manual AHL, Edition 09/2016
www.ankersmidsampling.com
Page 8 of 18
© Reproduction of this document or its content
is not allowed without permission of Ankersmid.
Technical Details Subject to Change