3M 701 Руководство по эксплуатации - Страница 7
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Измерительные приборы 3M 701. 3M 701 9 страниц. Megohmmeter
Также для 3M 701: Руководство по эксплуатации (8 страниц), Процедура калибровки (7 страниц)
B. Before testing new floor mats or newly installed floors,
clean mats/floors per manufacturer's recommendations.
For testing of floor finishes or monitoring of existing
floor materials, test in an as-is condition.
C. Perform tests at ambient humidity.
D. Place the Instrument and test weight at the desired test
location.
E. Connect one lead of the meter to ground with supplied
clip and the other lead to the test weight.
F. Set meter to 100V. Place test weight on the surface of
the material being tested.
G. Push test button and record the resistance after the
measurement has stabilized or after 15 seconds. Release
test button.
H. Repeat the procedure placing the test weight on the
surface at different locations.
I. Perform a minimum of five tests per contiguous floor
surface material or a minimum of five tests per 5,000
2
square feet (464,5 m
) of floor material, whichever is
greater. A minimum of three of the five tests should be
conducted in those areas that are subject to wear or have
chemical or water spillage or are visibly dirty.
TEST PROCEDURE FOR RESISTANCE
POINT TO POINT
A. Complete BATTERY TEST and CONTINUITY TEST.
If required clean electrodes as described in Test Weight
Cleaning section below.
B. Before testing new floor mats or newly installed floors,
clean mats/floors per manufacturer's recommendations.
For testing of floor finishes or monitoring of existing
floor materials, test in an as-is condition.
C. Perform tests at ambient humidity.
D. Place the Instrument and test weight at the desired test
location.
E. Connect test leads of the meter to the test weights.
F. Set meter to 100V. Place test weights three feet apart on
the surface of the material being tested.
G. Push test button and record the resistance after the
measurement has stabilized or after 15 seconds. Release
test button.
H. Repeat the procedure placing the test weights three feet
apart on the surface at different locations.
I. Perform a minimum of five tests per contiguous floor
surface material or a minimum of five tests per 5,000
2
square feet (464,5 m
) of floor material, whichever is
greater. A minimum of three of the five tests should be
conducted in those areas that are subject to wear or have
chemical or water spillage or are visibly dirty.
MAINTENANCE
GENERAL
This section details the maintenance which may be
performed by the user. Performance by the user of any
maintenance not specified here will void the Instrument
warranty. If it is determined that maintenance beyond the
scope of this section is required, contact your local 3M sales
representative or call 3M Technical Service (Electronic
Handling & Protection Division) at 512-984-3200.
Figure 6-1: Battery Replacement
BATTERY REPLACEMENT
Before attempting to replace either battery, place the main
selector switch in the OFF position.
The circuitry enclosed in the Model 701 produces high
voltages. Make sure that the main selector switch is in the OFF
position before removing the back cover.
A. To open the back cover, remove the screw located in the
center of the back cover.
B. The batteries are held in place by a metal bracket at the
top of the Instrument. To release this bracket, turn the
screw located in the center of the bracket counter
clockwise until the bracket swings free. The batteries
will now slide out.
C. Install new batteries as shown in Figure 6-1.
Improper battery installation will damage this Instrument.
D. Replace bracket and tighten bracket screw. Replace back
cover and cover screw.
TEST WEIGHT CLEANING
The test probes included in this kit are heavy. Exercise care in
handling.
After a period of use, the conductive rubber pads on the test
weights may become soiled, causing the weight to fail the
CONTINUITY TEST. To clean the surface of the
conductive pad, use a 70% Isopropyl alcohol/water mixture
on a clean low-linting cloth. Allow surface to "air dry" 15
minutes before use.
ZERO ADJUSTMENT
On occasion, due to handling, vibration, or other causes, the
pointer on the Model 701 may need adjustment. To zero the
pointer, turn the main selector switch to the OFF position.
Place the Instrument on a level stable surface and turn the
mechanical zero adjust screw until the pointer is over the
left-most mark on the OHMS scale.
5