Calibration and Earth Ground Testing:
It is recommended that the top speed, ground continuity and line leakage be
tested every two years for continued safe operation. Contact Drucker Diag-
nostics for further information or testing availability.
Transportation:
During transportation, accessories can become dislodged from the rotor and
fall into the rotor chamber where they can remain unnoticed.
Before transporting the centrifuge in a vehicle; make certain that the centri-
fuge is secured sufficiently to prevent shifting sliding or rolling, and that there
are no tube holders or accessories in the rotor or rotor chamber.
Notes on Repaired Centrifuges:
For verification of the safe state of the centrifuge following factory service or
repair, refer to the date on the calibration label (located on the back of the
centrifuge). This is the date that the repaired centrifuge was last factory test-
ed and calibrated. For additional servicing information and technical support,
contact Drucker Diagnostics or your authorized distributer.
Disposal:
This device is labeled with the WEEE symbol (a crossed out "waste bin"). The
final user should not discard this product along with other household waste. It
should be collected and treated separately.

Replacement Parts:

Part No.
7724037
7751068
7786047
7760016
7735054
7751079
7714101
7714103
7724071
7732018
7713079
7713032
02-002-1-0011
Description
Foot, rubber
Switch, lid safety
Rotor, six-place, fixed angle
Power Cord
Motor
Fuse
Pawl, latch, lid
Knob, latch, lid
Hinge, friction
Seal, lid gasket
Shield for 75/100 mm tubes
125 mm tube holders
Lid Assembly