Place the base assembly onto the guard bowl assembly with the guard bowl splice
positioned as shown (fig 9).
•
Feed the motor wire harness into the base passthrough so that the wire harness is
not pinched between the base and the guard bowl (fig 10).
Fig 11
•
Secure the base with six #8 screws torqued to 10 in-lbs. (fig 11).
•
Install the exhaust cover and secure with three #8 screws torqued to 10 in-lbs. (fig
12).
DRUCKER DIAGNOSTICS
HORIZON 12 SERVICE MANUAL
SPLICE
THIS SIDE
Fig 10
Fig 12
SM015
REV A