DS Maref DVT-2600 Руководство по эксплуатации - Страница 11
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Медицинское оборудование DS Maref DVT-2600. DS Maref DVT-2600 16 страниц.
11. Controller Maintenance and test methods
11-1. Test mode Set-up
When you press 'LEG' and 'MODE' buttons for 2 seconds at the same time, test method mode
including 10 test programs starts. Whenever you press the 'mode' button, the test number is
changed from 1 to 10 and when you press the 'START', you can change the selected value or do
the test. If you want to finish test method, please press 'START' for 3 seconds.
11-2. Test mode table
MODE 1
Check the program version
MODE 2
To remove the lymph mode
MODE 3
Check voltage for charging battery
MODE 4
Check battery voltage
MODE 5
Check pressure sensor or calibration
MODE 6
LCD Test
MODE 7
Solenoid valve test
MODE 8
Pump tset
MODE 9
Battery test
MODE 10
Sleeve leakage test
11-3. Test method
No
MODE
1
MODE 1
2
MODE 2
3
MODE 3
DISPLAY
- The version indication of controller's
software
- Left : MAIN Version
- Right : POWER Version
LYMPH MODE can be deleted if DVT
MODE is only used
- When you press "START". You can
turn on out the 'LYMPH' mode.
- Check the supply voltage of controller
- If the supply voltage of controller is
Description
minimum
average
14.5V
15.0V
below the minimum voltage, the
recharge may not be completed.
maximum
15.5V