Anritsu Company ShockLine MS46122B Series Краткое руководство по эксплуатации - Страница 9
Просмотреть онлайн или скачать pdf Краткое руководство по эксплуатации для Измерительные приборы Anritsu Company ShockLine MS46122B Series. Anritsu Company ShockLine MS46122B Series 20 страниц. Vector network analyzers verification kits and performance verification software
Setup the VNA and Verification Performance Software communication
The Performance Verification Software program must be set-up to communicate with the VNA. To locate the
VNA Network Address, on the VNA select: Main Menu | System | Network Interface.
Refer to
Figure
8. To set-up the VNA/Performance Verification Software (PVS) program communication:
1. From the PVS Select Interface pull down menu:
a. For MS46122B, select TcpipSocket.
b. For MS46322B, MS46522B, and MS46524B, select Tcpip.
2. Locate the PVS Address field.
3. Enter the IP Address as is displayed from the VNA Network Interface IP Address.
4. Enter the TCP Port number from the Network Interface menu into the Port field.
ShockLine Software must be running when running PVS with a single controller on a MS46122B.
Note
When running the PVS for a MS46122B, enter local host address (127.0.0.1) into the PVS Address
field and enter 5001 into the Port field.
1
1. ShockLine VNA Address
2. Select Interface
3. Address Field
4. Port Field
Figure 8. Instrument Address
3-2300-610 Verification Kits/PVS QSG
5. Installing the ShockLine Performance Verification Software (PVS).
PN: 10410-00766 Rev. B
2
3
4
QSG-9