Anritsu Company ShockLine MS46322B Series Краткое руководство по эксплуатации - Страница 15
Просмотреть онлайн или скачать pdf Краткое руководство по эксплуатации для Измерительные приборы Anritsu Company ShockLine MS46322B Series. Anritsu Company ShockLine MS46322B Series 20 страниц. Vector network analyzers verification kits and performance verification software
Test(s) that appear grayed-out indicate that not all equipment required those tests have been
entered in the Station Asset Utility list. See
Note
display the test equipment that is not available. To enter the asset of the missing test equipment, go
Section "To Add More Components:" on page
to
Figure 17. Grayed Test Sequences
Run the verification test
1. Select Run Test.
• A status light indicator appears next to the Run Test/Abort Test buttons. The light displays status
of the testing process for each test. Green is pass. Red is fail. Yellow is action required.
Figure 18. Light Test Status Indicator
• As each test is selected and run, the Activity Log will display the testing activity. See
above.
2. Continue to run all the tests until the Performance Verification Testing is complete.
3-2300-610 Verification Kits/PVS QSG
7. Starting Performance Verification Testing
Figure
17. Hover the mouse over the grayed test to
QSG-11.
PN: 10410-00766 Rev. B
Figure 16
QSG-15