4.11 Starting and Shutting Down the TRIOS System

Please follow the steps below to start TRIOS system for the first time:
Step 1: Turn the PC on.
Start 3Shape software.
Step 2: Log on using a default TRIOS Operator.
Step 3: Configure and use your TRIOS system.
Follow on-screen instructions for connecting to the wireless TRIOS scanner.
When using the TRIOS system, the user is guided by messages shown in the TRIOS software.
The TRIOS scanner has an LED ring which indicates the status of the scanner:
A rotating blue light indicates that the TRIOS scanner
is waiting for the TRIOS software to connect to it.
Three blue blinks indicate that the TRIOS scanner is
searching for a scanner application to connect to.
A constant blue light indicates that the TRIOS
scanner is connected to the system, but no
scanner tip is mounted.
A rotating green light indicates that the TRIOS
scanner is heating the TRIOS scanner tip.
A constant green light indicates that the TRIOS
scanner is ready to scan.
A rotating white light indicates that the TRIOS
scanner is waiting for scan data.
A constant white light indicates that the TRIOS
scanner is scanning. The light intensity decreases
if the TRIOS scanner loses registration.
Solid yellow indicates that the battery is low.
Scanner operation
The scanner has two buttons for operation; the button closests to the tip controls scanning
operation, the button farthest from the tip controls navigation.
Power saving
The TRIOS scanner has integrated power control, and will go into sleep mode when not used,
unless the software is on the Scan page.

4.12 Starting TRIOS Online Help

The TRIOS online help can be started from the TRIOS software by clicking on the Help
button and choosing User Manual.

4.13 Storage, Operating and Transport Conditions

Operating Conditions
Ambient temperature range: +15 – +26 °C (+59  – +78.8 °F)
Relative humidity: 10 – 85 % (non-condensing)
Atmospheric pressure: 800 – 1100 hPa
The TRIOS scanner should be used in an environment maintaining a
consistent room temperature.
Storage Conditions
Ambient temperature range: -10 – +50 °C (+14 – +122 °F)
Relative humidity: 10 – 85 % (non-condensing)
Atmospheric pressure: 800 – 1100 hPa
Transport Conditions
Ambient temperature range: -10 – +50 °C (+14 – +122 °F)
Relative humidity: 10 – 85 % (Non-condensing)
Atmospheric pressure: 620 – 1200 hPa

4.14 Technical Specifications

Light source
Do not view directly with optical instruments, such as spectacle
specifications
microscopes. The product complies with EN (IEC) 62471
(Photobiological safety of lamps and lamp systems and UL-standard
US code of federal regulations, 21 CF 1040.10 and 1040.11.
Sterilization
Sterilize by autoclave (see section 7 for autoclave specifications)
Power input
C2BP: Battery charger
Power input: 100-240 VAC 0.6 A, 50/60 Hz
Protection class: IP20
Insulation class Class II
L1-2 scanner. Wireless, battery powered
3Shape TB-04 Battery:
3,6 V, 3300 mAh, 12 Wh
Wireless
The TRIOS scanner contains an IEEE 802.11ac module
Wireless
Maximum 100 mW
Transmit Power
Wireless
2.401-2.483 GHz
Frequency Band
5.15-5.725 GHz (the actual frequencies are dependent on local
regulations and the configuration of the product.)
IT Network equipment
802.11ac (5GHz)
Network
characteristics
Capacity for at least 40 Mbps per scanner.
Network
Bi-directional traffic permitted between TRIOS scanner and PC.
configuration
mDNS enabled (required for scanner discovery).
Redirection of traffic must be disabled.
Channel width set to at least VHT40 (40MHz).
Ports used:
• TCP ports (Scanner): 20-21, 80 and 23795
• UDP Ports (Scanner): 58218 and 58220-58230
• UDP Ports (PC and Router): 67 (DHCP), 5353 (mDNS)
Security
WPA2-PSK authentication
specification
IMPORTANT NOTICE
If you connect your TRIOS intraoral scanner system to your IT Network
equipment, it is strongly recommended to apply network security
measures and to analyse, evaluate, and control the IT network.

4.15 Electromagnetic Compability

The TRIOS system has been tested according Medical electrical equipment – Part 1-2:
General requirements for basic safety and essential performance – Collateral Standard:
Electromagnetic disturbances – Requirements and tests, EN/(IEC) 60601-1-2 ed. 4.0, 2015.
WARNING
Medical electrical equipment needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC
information provided in this document.
IMPORTANT NOTICE
Portable and mobile RF communications equipment can affect medical
electrical equipment.