Freedom Innovations iQ Руководство по настройке - Страница 8
Просмотреть онлайн или скачать pdf Руководство по настройке для Медицинское оборудование Freedom Innovations iQ. Freedom Innovations iQ 16 страниц. Vacuum device
Также для Freedom Innovations iQ: Руководство пользователя (10 страниц), Руководство по эксплуатации (11 страниц)
Installation:
1. Remove the front cover and clamp (Image A).
2. Attach 1/8" tubing with filter to 90˚ barb in the socket.
3. Connect the 1/16" tubing to the filter assembly and route
it to the proximal barb on the iQ Vacuum pump. Cut the
tubing to length as necessary.
4. Select a rubber clamp bushing (Image C) to ensure the iQ
Vacuum unit fits snuggly against the pylon.
5. Position the unit as high as possible on the posterior side of
the pylon. Make sure there is no contact with the socket or
the foot when mounted. If needed, you may mount the iQ
Vacuum unit on any side of prosthesis for a better fit.
6. Attach the anterior clamp with the two clamp bolts (Image
B) and tighten with a torque wrench to 5Nm (4 ft-lb). Check
that the unit is held firmly and does not slip. You may have
to choose a different thickness of rubber clamp bushing if it
is not secure, or if you cannot tighten the clamp fully.
7. Ensure the tubing passes through the proximal relief cutout
in the housing (Image E). Then attach anterior cover (Image
A)
Note: Ensure there are no sharp turns or kinks in the tubing.
E
D
C
B
A