CONTEC MEDICAL SYSTEMS CO.,LTD

Notice items:

1) pulse oximeter probe placement, the position without ductus arteriosus, BP cuff and vein input pipe is top-priority.
2) If the pulse oximeter probe can't monitor the state of pulsation, it shows that the position of probe is improper, or the position is too
thick, too thin or having too deep pigment to reach a proper translucidus effect. If above things has happened, place the probe again or
select probe of other type.
3) This pulse oximeter probe should be applied to the special medical equipment. Operator is responsible to check the compatibility
Incompatible fittings or device will influence the measuring result.
4) The disposal of scrap instrument and its accessories and packing (including battery, plastic bags, foams and paper boxes) should
follow the local laws and regulations.
Maintenance/cleaning/disinfection:
1) Check if the product is undamaged and clean before using.
2) This product is not allow to use disinfection liquid for disinfection, this probe belong to one-off products.
Note: Don't immerse the product in the liquid, and don't expose it under the strong ultra-violet radiation
Service life:
Suggest this product use only once, don't use again.

Environment requirements:

Transport and storage
1)
Temperature: -10℃~+40℃
2)
Humidity: less than 80%
3)
Pressure:500hPa~1060hPa
Operating
1)
Temperature: 10℃~+40℃
2)
Humidity: 30% ~ 75%
3)
Pressure: 700hPa~1060hPa
Statement:
1)
pulse oximeter probe needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided in User Manual and test report.
2)
Portable and mobile RF communications equipment can affect pulse oximeter probe.
Warning:
1) The use of cables other than those specified, with the exception of cables sold by CONTEC as replacement parts for internal
components, may result in increased emissions or decreased immunity of pulse oximeter probe.
2) pulse oximeter probe should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary,
the pulse oximeter probe should be observed to verify normal operation in the configuration in which it will be used.
3) Improper usage can result in inaccurate measurement.
4) Using it under too strong light will cause inaccurate measurement, in case of that, please set a opaque stuff around the probe to cut
light off.
5) You should move the probe to other position per 4 hours at least. Because the state of local skin can influence the ability of skin to
enduring probe, it is necessary to replace the position of probe according to the state of patient. Please do that when skin integrity
changes.
6) The dyestuff in blood vessel cab cause the inaccurate measurement.
7) The performance of pulse oximeter probe is influenced by movement easily, so it is not suitable for active patient to use it.
8) Don't fix the probe with belt or bundle it tightly, because the vein pulsation can cause inaccurate SpO
9) Same as other medical equipment, the cable should be set properly to avoid enlacing or asphyxiate patient.
10) Don't use it in the process of MRI scan, because the conductor current may burn the skin of patient, moreover, the probe will
influence MRI image and MRI set will also influence the accuracy of SpO
11) Don't change the product at will, otherwise the capability or accuracy of product will be influenced.
12) The probe is not intended for use during patient transport outside the healthcare facility.
13) DO NOT use the probe while the patient is being scanned by MRI or CT.
measurement.
2
measurement.
2
2