DARAY V406 Руководство пользователя - Страница 2

Просмотреть онлайн или скачать pdf Руководство пользователя для Медицинское оборудование DARAY V406. DARAY V406 2 страницы.

DARAY V406 Руководство пользователя
V406 User Manual
Maintenance & Storage
Cleaning
• U se a soft cloth and medical alcohol to clean the
device, and allow to air dry.
Maintenance
• C leanthedeviceregularlyasperabove
• K eepthebatterychargedandrechargewhen
prompted by the V406
•Conditionthebatteryevery6months,especially
during periods when the device is rarely used.
Storage
• T heV406isbeststoredinoriginalpackaging
• R emove the battery during long storage periods
• D onotstorewithcorrosiveorexplosivegases
•Storewithplentyofventilation
• S toragetemperaturerange:-40to60°C

Troubleshooting

Problem
Cause
Abnormal or eratic
Thepatientsfingerisnotcorrectlypositioned
SpO2/PRreadings
ThepatientsSpO2/PRistoolowtobedetected
Thepatientismoving/shaking
The patient is wearing nail varnish
V406 is not
The battery has little or no charge
powering on
The battery is not installed or is not correctly
installed
The V406 is faulty
The display
The battery has little or no charge
powers off
suddenly
The V406 is set to power off after 5 seconds when
not in use
The unit can not
The battery has little or no charge
be used for long
periods of time
The battery is damaged
DARAY Ltd. Marquis Drive, Moira, Derbyshire DE12 6EJ
Tel: +44 (0)333 321 0971
DOCUMENTNo.QAM.SUG.V406.1012.1
Quick Setup Guide
Specification
Sp02
Range
0-100%
(1%Resolution)
Accuracy
±2%at70-100%
PulseRate
Range
30-250BPM
(1BPMResolution)
Accuracy
±2BPMor±2%
Dimensions 46(L)x40(W)x29(H)mm
Weight
Approx.35g(inc.battery)
Operating Environment
Temp.
10°C-40°C
Humidity
30%-75%
Pressure
700hPa1060hPa
InfaredWorkingElements
Solution
Removethepatientsfingerandreinsert
Removepatientsfingerandattemptreadingagain.
askpatienttoremainasstillaspossible
Removeanynailpolishfromthepatientsfinger
Charge the battery for at least 8 hours using the
charger provided
Checkthebatteryisinstalledandthatthepolarity
is correct
Return to DARAY for inspection and repair
Charge the battery for at least 8 hours using the
charger provided
Normal operation
Charge the battery for at least 8 hours using the
charger provided
Contact DARAY for replacement
Fax: +44 (0)333 321 0973
DARAY Ltd. registered in England no. 05276678
www.daray.com