Apogee Instruments MQ-620 Руководство пользователя - Страница 11

Просмотреть онлайн или скачать pdf Руководство пользователя для Измерительные приборы Apogee Instruments MQ-620. Apogee Instruments MQ-620 15 страниц. Quantum meter

Apogee Instruments MQ-620 Руководство пользователя
Moisture or debris on the diffuser is a common cause of low readings. The sensor has a domed diffuser and
housing for improved self-cleaning from rainfall, but materials can accumulate on the diffuser (e.g., dust during
periods of low rainfall, salt deposits from evaporation of sea spray or sprinkler irrigation water) and partially block
the optical path. Dust or organic deposits are best removed using water or window cleaner and a soft cloth or
cotton swab. Salt deposits should be dissolved with vinegar and removed with a soft cloth or cotton swab. Never
use an abrasive material or cleaner on the diffuser.
Although Apogee sensors are very stable, nominal accuracy drift is normal for all research-grade sensors. To
ensure maximum accuracy, we generally recommend sensors are sent in for recalibration every two years,
although you can often wait longer according to your particular tolerances.
Steps to Replace a Handheld Meter Battery
1. Use a phillips head screw driver to remove the screw from the battery cover.
2. Remove the battery cover by slightly lifting and sliding the outer edge of the cover away from the meter.
3. Use your thumb to slide the battery out of the battery holder.
a.
If the battery is difficult to move, turn the meter on its side so that the opening for the batter is
facing downward and tap the meter downward against an open palm to dislodge the battery
enough that it can be removed as described in step 3.
4.
To place the battery back in, simply slide it back into the battery holder with the flat side of the battery
facing up.