DURRIDGE RAD7 Руководство пользователя - Страница 13
Просмотреть онлайн или скачать pdf Руководство пользователя для Измерительные приборы DURRIDGE RAD7. DURRIDGE RAD7 15 страниц. Passive, slow response radon in water accessory
Также для DURRIDGE RAD7: Руководство по подключению Bluetooth (4 страниц), Руководство по подключению (4 страниц), Руководство пользователя (7 страниц)
5 CARE, MAINTENANCE, AND TROUBLESHOOTING
5.1 Water Catastrophe
If water ever enters the RAD7, or if the RAD7 ever
goes swimming in the water, it will probably cease to
operate and immediate steps should be taken to
minimize the impact on the instrument.
Keep the RAD7 upright. This will prevent water from
touching the detector, which is close to the face plate
at the top of the dome. Put a piece of tubing on the
RAD7 outlet with the other end in a sink. Use the
RAD7 pump if it still works or, otherwise, an external
pump into the inlet, to blow air through the
instrument. When water ceases to be blown out of
the outlet, put desiccant upstream of the RAD7 to dry
out the air path. When the air path is fully dry (after
dry air has been blown through it for approximately
one hour), remove the face plate from the case, empty
the water out of the case and blow dry the case and
the RAD7 electronics.
Once there is no visible water in or on the
instrument, it can be put in an oven at 50°C for a few
hours to dry out completely. Additionally, desiccated
air can be passed through the air path until the air
leaving the RAD7 drops below 10% RH. After this
treatment further corrosion will be prevented, and
the RAD7 will boot once more and you can use the
internal RH sensor to measure how dry the air path
is. At this point the instrument should be returned to
DURRIDGE for service.
5.2 RAD7 Care
Water, particularly salt water, is hostile to electronic
instruments. Please keep the RAD7 in a relatively
clean and dry environment. One way is to enclose the
instrument in a large, transparent plastic bag; see
Section 1.2.3. Should it ever be seriously splashed
with salt water, please follow the instructions in
Section 7.1, above.
As a preventive measure, plastic cling wrap can be
placed over the RAD7 face plate and down the sides
of the RAD7. Push it down around the hose
connections, push the power and RS232 plugs into
their sockets, and push the lid onto its hinges. The
wrap will make the RAD7 almost watertight. If it
tears it can be easily replaced at any time. The
instrument should, in any case, be returned every
year for recalibration.
Section 5
Care, Maintenance, and Troubleshooting
It is useful to look at a cumulative spectrum
periodically. This may be obtained by having the
printer on and allowing the RAD7 to complete a run.
The "Recycle" number may be set to the current cycle
Setup, Recycle, NN [ENTER]
number (
the RAD7 reaches the end of the current cycle it will
then print out the end of run summary including the
cumulative spectrum. Look to see that the peaks are
clean and in their normal positions.
5.3 Exchanger Care
There is a potential for biofouling to occur in long-
term monitoring applications. The exchanger should
be kept as clean as possible in the circumstances. The
Water Probe should therefore be inspected
periodically to ensure that there is good contact
between the radon exchange tubing and the
surrounding water, and no clogging with weeds,
algae, etc.
5.4 Desiccant Regeneration
Please see the RAD7 manual for information on the
care and regeneration of the desiccant. Regenerated
desiccant, after a few regenerations, loses most of its
indicating ability (due, we believe, to migration of the
cobalt chloride to the interior of the calcium sulphate
crystals). One way to 'indicate' the status is, every
time you refill the laboratory drying unit with
regenerated desiccant, you first add half an inch or so
of new, blue desiccant, out of the jar. This way, you
can always tell if the unit is still working, as the new
desiccant will only turn pink when the rest of the
desiccant, upstream, has become hydrated.
5.5 Water Probe Troubleshooting
5.5.1 Air Path Integrity
When drawing a sample from a remote location, air
path integrity is essential to prevent dilution of the
sample with ambient air. Always make sure that there
are no loose connections or leaky fittings (such as the
screw cap of the Laboratory Drying Unit) in the air
loop, particularly upstream of the RAD7. In the event
of unexpectedly low radon values, check the air path
for integrity.
). When
13