ADE M10010 Руководство по эксплуатации - Страница 4

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Весы ADE M10010. ADE M10010 12 страниц. Electronic baby scale, class iii approved

ADE M10010 Руководство по эксплуатации
Your ADE Electronic Baby Scale is a class III approved scale which has been manufactured for
medical use in hospitals, by doctors or medical specialists.
The scale is a precision instrument.
The weighing result is used as an essential basis for medical application decided by doctors or
medical personnel in hospitals, nursing homes, rehabilitation- and dialysis centres.
Make sure the baby is only weighed by trained personnel and is never left unattended.
In order to achieve the best possible results kindly read and follow the Operating Manual in all
respects.

Check carton contents

-
Baby scale
-
Baby tray
-
Mains adapter
-
Operating Manual including Certificate of Conformity

Power Supply of the scale

By mains adapter (9V 800mA) and rechargeable batteries (power-pack).
-
The battery pack is already installed in the base of the scale.
-
Charge the power-pack by using the mains adapter and connect the jack firmly.
-
Complete first charging of power-pack needs 24 hours.
-
The full battery life is approx. 100 hours.
-
If power-pack needs recharging there will be an arrow above the low battery symbol on the
display. Still the scale with operate for about 10 hours after which it will automatically switch
off to protect the battery.
-
You can also identify status of battery charging by the colour of the LED control light placed in
the left hand corner above the display.
-
When the scale is plugged into mains the power pack will be recharged automatically.
-
If LED is green the batteries have full charge. If it is red the battery is nearly discharged and
yellow indicates the battery is under charge.
-
Battery life becomes weaker if lifetime after charging becomes shorter. In this case contact
your local dealer or service partner.
-
Please make sure the local voltage supply corresponds to the indication printed on the
adapter.

Positioning of the scale

-
Place the scale on a level surface.
-
Make sure all four feet of the platform have direct contact to the surface.
-
The spirit level of the scale is an additional tool to make sure the scale is carefully positioned
and offers best results when the air bubble is centred.
-
The alignment of the scale must be checked and corrected every time the location of the scale
is changed.

Use of the scale where approved scales are imperative

The scale corresponds to standards and requirements of the European Directives 90/384/EEC and
93/42 EEC.
3

Installation and use of scale

050100-Rev000-M10010