Datel MAY600 PON OTDR Руководство - Страница 18
Просмотреть онлайн или скачать pdf Руководство для Измерительные приборы Datel MAY600 PON OTDR. Datel MAY600 PON OTDR 19 страниц.
The battery inside the instrument is a special lithium battery. In order to give full play to the performance of the battery,
please use the internal battery power supply when you start using the instrument. The first use of the battery needs to be
depleted, and then charging the battery, the first charging time should be no less than 10 hours. Battery charging
temperature range is 0°C~50°C. For your safety, the charge will terminate automatically when the charging temperature is
too high. When the instrument is idle for more than two months, it should be charged in time to maintain the battery power.
Please don't take out the battery without permission. Please do not let the battery close to the fire source or strong heat.
The battery should be removed when the instrument is stored for a long time. The storage temperature range of the
battery is -20°C~45°C.
Instrument screen cleaning and touch calibration
The display of the instrument is a 5.8 inch color LCD with touch screen. Please do not use sharp objects to click on the
screen, otherwise it may lead to damage to the LCD screen. You can clean the LCD screen using soft fabric, but do not
use organic solvent.
Instrument calibration
It is recommended that the instrument be calibrated once every two years. Please contact the supplier for specific items.
Common faults and troubleshooting
Common faults
Instrument can not start normally
Instrument can not charge properly
Inaccuracy test results
Noise too high of test curve
Saturation of curve front end
Slow down of curve front end
Unable to measurement
the end of fiber
Event analysis error
Optical fiber length inaccuracy
Dirección: Calle México 1554 Edif. Chuquiago MZ-2
Teléfono/Fax: (2)2-231-1125 / 211-4820
E- E-mail:[email protected]
Failure cause
Battery low power
Ambient temperature is too high or low
Poor battery connect
Instrument problem
Incorrect parameter setting
Optical fiber end contamination
Connector type mismatch
Optical port damage
Connector type mismatch
Pulse width too small
Incorrect attenuation
Pulse width too large
Incorrect attenuation
Optical fiber end contamination
Optical port damage
Connector types mismatch
Range too small
Pulse width too small
Incorrect test conditions
Incorrect test conditions
Incorrect fiber refractive index
Incorrect optical cable correction
Cel.: 71521900 / 71535025
Pagina web: www.datelsrl.com
Troubleshooting
Battery charging
Charge the battery in the 0℃~45℃
Check battery connectors
Contact supplier
Resetting parameters
Clean the end face of fiber
Replacing matching connector
Replacing fiber connector
Replacing matching connector
Increasing pulse width
Modify attenuation
Reducing pulse width
Modify attenuation
Clean the end face of fiber
Replacing fiber connector
Replacing matching connector
Increasing test range
Increasing pulse width
Modify test conditions
Modify test conditions
Modify fiber refractive index
Modify optical cable correction
18