4titude 4s2 Руководство по эксплуатации - Страница 13
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Экономия продуктов 4titude 4s2. 4titude 4s2 14 страниц. Variable temperature thermal sealer
4s2 Variable Temperature Thermal Sealer Service Manual
3.5 Left-to-Right Adjustment
1.
Remove the cover.
2.
Locate the lower M4 screw underneath the platen that receives the PCR / microplate (see
Figure 10). If the unit is under-sealing on the right, loosen the screw and tilt the lift to the left.
If the unit is under-sealing on the left, loosen the screw and tilt the lift to the right.
3.
Re-tighten the screw.
4titude Ltd, The North Barn, Surrey Hills Business Park, Damphurst Lane, Wotton, Dorking, Surrey, RH5 6QT, UK
Tel: +44 (0) 1306 884885 Fax: +44 (0) 1306 884886 [email protected] www.4ti.co.uk
Figure 12 - Left-to-Right Adjustment
Page 12