activecare medical Spirit Руководство пользователя - Страница 12
Просмотреть онлайн или скачать pdf Руководство пользователя для Инвалидное кресло activecare medical Spirit. activecare medical Spirit 19 страниц. Folding power wheelchair
6. Battery Charger
1
6
7
8
1 . Output fuse of anti-polarity counter
2 . Charge cable
3 . Power switch
4 . 115V/230V switch
5 . Socket of power cable
6 . Radiating fan
7 . Red light: the power input indicator
8 . Yellow light: being charged
9 . Green light: charging finished
6.1. Specifications:
Mode
Specifications
Input Voltage
Output Voltage
Output Current
Max. Output Voltage
Protection Against Over-Voltage
Protection Against Overload
Protection Against Short Circuit
Environmental Temperature of use -10°~ 50°
Environmental Temperature of
Preservation
Dimensions
Weight
Approvals
2
3
115V/230V Option Switch
115V(95V-132V)/230V(180V-264V)
24V
3A~6A Continuous
31V
Above 32.5V-Automatic Shut-off
Above 8A-Automatic Shut-off
Output Fuse Severs of Shut-off Automatically
-25°~ 85°
180mm(L) x 93mm (W) x 53mm (H)
0.83kg
UL, CSA, CE, TUV, JAPAN
4
5
HP 8204B