Selecting the 'Allow Stop & Restart' option allows the button to remain active and control recording
stopping and starting. This is only recommended for laboratory testing.
'Immediately on Disconnect' starts recording as soon as the device is removed from the cradle.
The LEDs are inactive in this mode and the button will not interrupt recording, but instead will be
used as an event marker only, when required.
'At Future Time' allows the operator to choose a start time up to 2 weeks in the future. Recording
will start automatically at this point. The LEDs are inactive in this mode and the button will not
interrupt recording, but instead will be used as an event marker only, if required.
iv). Subject info
Once the date of birth is selected, the age will be calculated automatically. Enter the height in cm
and weight in kg. The height in feet/inches, weight in stones/pounds and BMI will then be calculated
automatically.
We do not recommend entering personal data, such as date of birth, as this can be entered post
data collection if required. It does not directly affect any data collected and means there is no
Personally Identifying Information stored on the device.
v). Configuration
Select which device(s) are to be configured. A tick box will appear next to the relevant device serial
number. At least one device must be selected. It can be useful to configure multiple devices together
when a trial subject will wear several GENEActiv devices on different body locations.
When 'Erase & Configure' is clicked, a pop-up message will automatically appear to warn that
continuing will erase all previously stored data and to confirm the recording start mode. There will also
be warning pop-ups if any mandatory information is missing, or if the battery charge is below 80%. If
the warning message appears due to the battery being less than 80% do not deploy the device.
Device configuration will take about 10 seconds and a pop-up message will automatically appear to
confirm that it has been set up successfully and is ready for deployment. The device will record for the
given data collection period. There is no need for any additional charging (this should be avoided due to
the important note below).
IMPORTANT: Once the configuration of the device is complete, it is important that the device
is not returned to the charger cradle at any point, until after data collection. If the device is re-
entered into the cradle for any reason, it will stop data recording and must be reconfigured to start
the data collection process.