QUICK MENU
QUICK MENU consists of usually used parameters for quickly parameter setting.
1) Parameters of QUICK MENU
Page
Parameter
Description
CH
Select channel for the QUICK MENU[1] setting.
TYPE
Input type
QUICK
DOT
Decimal point
MENU [1]
UNIT
Display/Temperature unit [INPUT SETUP]-[Display/Temp Unit]
PEN COLOR
Pen color
CH
Select channel for the QUICK MENU[2] setting.
Low-limit input value or
LOW RANGE
graph scale value
QUICK
High-limit input value or
MENU [2]
HIGH RANGE
graph scale value
LOW SCALE
Low-limit scale value
HIGH SCALE High-limit scale value
PRINT MODE Record mode
PRINT SPEED Standard record speed
QUICK
PRINT MEMO Digital memo period
MENU [3]
BACK LIGHT
LCD backlight
LCD ON/OFF LCD backlight On/Off
USB REC
Memory save
USB COPY
Call USB COPY window
QUICK
MENU [4]
UPGRADE
Call upgrade window
CANCEL
Cancel the settings
SAVE
Save the setting of QUICK MENU[1] to [4].
※Refer to "KRN100 user manual".
Parameters
RUN mode
※ Dot parameters may or may not appear, depending on the other parameter setting.
3 sec.
INPUT SETUP
ALARM SETUP
(input setting)
(alarm setting)
Select UI Card
Select UI Channel
(universal input
(universal input
card select )
select)
Input Set Copy
Alarm Set Copy
(input parameter
(copy alarm
copy)
parameter)
LCD/Paper Record
Alarm
type
(display and record
(alarm operation
measuring value)
mode)
Alarm
Ref Channel
Pen Color
(alarm
reference
(record color)
channel)
Record Zone
Alarm
Option
(record zone)
(alarm
option)
Tag Name
Alarm
Value
(channel name)
(alarm
SV)
Input Type
Alarm
Hysteresis
(input
specifications)
(alarm
hysteresis)
OK,
Alarm
ON/OFF Delay
Range/Scale Point
Cancel
(alarm
output ON/OFF
(decimal point)
delay time)
Display/Temp Unit
Alarm
Alarm No
(display/temp.
(alarm
output
unit)
alarm number)
High/Low Range
Selection Alarm Card
& Graph Scale
(Alarm output card
(high/low input value
select)
and graph scale)
Low Scale/
Alarm-
Status
High Scale
(relay and transistor
(high/low scale value)
output method
Special Func
(special function)
Two Unit
(display vacuum,
static pressure)
Ref Channel
(reference
channel)
Input Bias
(error correction)
Span
(gradient
adjustment)
Record Method
(data storage
method)
Filter Type
(input digital filter)
Filter Counter
(number of digital
filters)
Burnout Action
(display setting
for break)
Operation
Trend graph
Screen
→ Bar graph
display
→ Digital number
QUICK
Usually used
MENU
parameters of
parameter setting group
are reorganize as
QUICK MENU.
Each parameter setting
Parameter
is linked.
setting
group
2) QUICK MENU Setting
①
Linked parameters
ᆞE.g.) In case of CH1, recording as input type=TC-U, low-limit input value=300, standrad record speed= 240mm/h
[INPUT SETUP]-[Input Type]
①
[INPUT SETUP]-[Range/Scale Point]
[INPUT SETUP]-[Pen Color]
[INPUT SETUP]-[Low Range] or
[INPUT SETUP]-[Low Graph Scale]
[INPUT SETUP]-[High Range] or
[INPUT SETUP]-[High Graph Scale]
[INPUT SETUP]-[Low Scale]
[INPUT SETUP]-[High Scale]
[RECORD SETUP]-[Record Mode]
[RECORD SETUP]-[Standard Speed]
[RECORD SETUP]-[Memo Period]
④
[SYSTEM SETUP]-[Backlight]
[SYSTEM SETUP]-[Backlight On/Off]
[FILE/MEMORY SETUP]-
[USB LogData Save]
[FILE/MEMORY SETUP]-
[USB Memory Copy/Move]
[USER/INFORMATION SETUP]-
[Firmware Upgrade]
DIGITAL INPUT
COMMUNICATION
SETUP
SETUP
(digital input setting)
(com. setting)
Select DI Card
Modbus Address
(digital input card
(com. address)
select )
DI-
Type
RS485 Port
(digital input
(RS485 com. use)
select)
DI-
Reset No
Baud Rate
(reset alarm
(com. speed)
number)
DI-
Status
Parity Bit
(operation status)
(com. parity bit)
Stop Bit
(com. stop bit)
Termination Set
(terminating
resistance)
Response Wait Time
(com. response
waiting time)
Protocol
(com. protocol)
RS485 Com/Write
(RS485 com.
write)
Ethernet Port
(Ethernet com.
use)
IP Address
(IP address)
Subnet Mask
(subnet mask)
Default Gateway
(default gateway)
Ethernet Com
Write
(Ethernet com. write)
USB Device Port
(USB com. use )
USB Com Write
(USB com. write)
Start/Stop
recording
Parameter
setting
information
output
3 sec.
RUN mode
Manual feed
(when stop
recording)
3 sec.
Digital memo
3 sec.
3 sec.
②
Press the
key once in RUN mode and it
enters to QUICK MENU.
QUICK MENU consists of usually used
parameters for quickly parameter setting.
②
Press the
key in RUN mode to enter
QUICK MENU.
⑤
Press the NEXT(
) key once and it moves
to QUICK MENU [3]. Press the SET(
) key
and set standard record speed [PRINT SPEED]
as 240mm/h.
RECORD SETUP
SYSTEM SETUP
(record setting)
(system setting)
Record Mode
Device Name
(record mode)
(device name)
Digital Print type
Date/Time
(1 line record CH during
(date/time)
numeric recording)
Standard Speed
Date Type
(standard record
(date type)
speed)
Option Speed
Summer Time
(option record
(summer time)
speed)
Memo Period
Summer Time
(digital memo
Period
cycle)
(summer time period)
Divide Zone
Alarm Sound
(record zone
(alarm sound)
division)
Standard Period
Sampling Rate
(standard record
(sampling cycle)
cycle)
Option Period
Log Speed
(option record
(save cycle)
cycle)
Listing Language
Backlight
(language for list
(LCD backlight
printing)
brightness)
Alarm Speed
Baklight On/Off
(alarm record
(LCD backlight ON
speed)
method)
Power On Status
(record status
when power ON)
Run Status
(list printing at
start recording)
List Out Option
(list record option
Zone Dot Line
Distance
(dot line for zone division)
CH Print Distance
(record interval for
each channel graph
Start Line Print
(start line when
starting record)
Range Print Time
(input range
record period)
Cautions during Use
1. Follow instructions in 'Cautions during Use'. Otherwise, it may cause unexpected accidents.
2. Install a surge absorber at each end of inductive load coil when controlling high-capacity power relay or inductive
load (e.g. magnet).
3. Check the polarity of the terminals before wiring the temperature sensor.
For RTD temperature sensor, wire it as 3-wire type, using cables in same thickness and length.
For thermocouple (CT) temperature sensor, use the designated compensation wire for extending wire.
Special
4. Keep away from high voltage lines or power lines to prevent inductive noise.
function for
In case installing power line and input signal line closely, use line filter or varistor at power line and shielded wire
record
at input signal line.
Do not use near the equipment which generates strong magnetic force or high frequency noise.
5. This unit may be used in the following environments.
①Indoors (in the environment condition rated in 'Specifications') ②Altitude max. 2,000m
③Pollution degree 2
③
Set the keys following the each parameter. Press
the NEXT(
) key and it moves to next page.
E.g ) When changing the temperature unit (℃→℉)
of CH1, press the SET(
) key.
③
Press the SET(
) key at QUICK MENU [1]
and below screen is displayed. Set input type
[TYPE] as TC-U by pressing SET(
)
,
keys and press the
key.
Press the NEXT(
) key once and it moves
to QUICK MENU [4]. Press the SAVE(
)
key to save the set ings of QUICK MENU [1] to
[4] and it returns to RUN mode.
RESERVATION
FILE/MEMORY
SETUP
SETUP
(file/memory setting)
(reservation setting)
Reservation
Load Set File
Type
(parameter setting
file open)
(reservation record)
Reservation Period
Save Set File
(reservation
(parameter setting
record period)
file save)
Reservation Time
Memory Status
(reservation
(memory capacity)
record time)
Memory Clear
(memory delete)
USB LogData
Save
(USB save function)
Memory Save Option
(memory save
option)
USB Memory
Copy/Move
(data move/copy)
④Installation category II
After completing the set ing, press the
SAVE(
) key at QUICK MENU[4] and save
the settings. It returns to RUN mode.
Press he NEXT(
) key once and it moves
to QUICK MENU [2]. Press the SET(
) key
using
keys to set
,
,
,
low-limit input range [LOW RANGE] as 300 and
press the
key.
USER/INFORMATION
RECORD BACKUP
SETUP
DATA
(user/info. setting)
backup data record setting
Record Backup
Password
(backup data
(password mode)
record)
Login Admin
Backup Data List
(admin. log in)
(backup data list)
Change Admin
Start Date and Time
(start time for data
Password
save)
(password change by admin.)
User Lock
End Date and
(user authority
Time
change)
(end time for data save)
Information
Backup Print Mode
(system info.
(backup data print
check)
mode)
Firmware Upgrade
Selection Print Mode
(firmware
(backup data recording
upgrade)
mode setting)