PREPARING THE NEBULIZER FOR USE
CAUTION
To prevent the risk of cross-infection, different persons should
never use the same nebulizer kit.
1. Place the device on a stable, sturdy, and flat horizontal surface
where you can easily reach the unit while you are seated.
Caution: place the device at least 4 inches (10 cm)
distance from wall.
2. Make sure that the unit is in the "Off" (O) position.
3. Insert the power plug into the electrical outlet.
4. Rotate the inhalation tube counterclockwise to remove
the inhalation tube from the medication tank.
5. Add the correct amount of prescribed medication to the
medication tank.
6. Turn the inhalation top clockwise until securely closed.
7. Attach the desired inhalation accessory.
ATTACHING IN INHALATION TUBE
1. Push the air plug on one end of the inhalation tube onto the
air connector on the back side of the compressor.
2. Push the air plug on the other end of the inhalation tube
onto the inhalation tube connector on the bottom of the
nebulizer kit.
USING THE DEVICE
1. Switch on the power.
2. Plug the inhalation tube into the
air plug on device.
3. Plug the other side of tube into the
vent port of bottom of nebulizer kit.
4. Add the medicine into the
nebulization tank.
5. Press the I/O button to turn on device
and begin treatment.
6. Press the I/O button to safely
terminate the operation of nebulizer.
7. Unplug the power cord from the
electrical outlet
after using the device.
CAUTION
The patient and operator must keep a distance from the device
of 20 cm to 50 cm (inches 8-20) in normal use.
USING THE MASK
Place the mask over the nose and mouth. Pull the elastic strap
over the head. Gently pull on the strap to securely hold the mask
over the nose and mouth. Inhale the medication. Exhale normally
through the mask.
CLEANING
CLEANING
Following the cleaning instructions after each use will prevent any
remaining medication in the bottle from drying, which can result in
the device not nebulizing effectively.
Rinse the nebulizer parts after each use.
1. Remove the inhalation accessory (mask, mouthpiece)
from the nebulizer kit.
2. Disconnect the inhalation tubing from the nebulizer.
3. Gently twist the inhalation top counterclockwise and lift
to separate the nebulizer into two sections.
4. Remove the baffle.
5. Discard remaining medication.
6. Rinse all the applied parts (nebulizer kit, the mouthpiece
and mask) in warm water.
7. Air dry only.
8. CLEANING THE DEVICE AND THE TUBE'S OUTER SURFACE
Only use a damp cloth.
9. Assemble the nebulizer and store the nebulizer kit in a dry,
sealed bag.
CAUTION: The nebulizer kit should be replaced every 6 months.
4
CARING FOR THE DEVICE
To keep your device in the best condition and protect the unit
from damage, follow these directions:
CLEANING THE COMPRESSOR
Clean the casing of the main unit by using a soft cloth
moistened with water or a mild detergent. Do not use abrasive
cleaners. Dry the casing immediately using a soft clean cloth.
CHANGE THE AIR FILTER
Change the air filter every 60 days even if the air filter does
not appear dirty. If the air filter appears dirty, or if water or
medication has spilled on the air filter, replace with a new air
filter immediately.
1. Pull the air filter cover to remove from the front side
of the compressor.
2. Remove the dirty filter by hand.
CAUTION: Do not attempt to wash or clean the air filter. Damp
air filters can cause blockages. Do not substitute cotton or any
other material for the air filter.
CAUTION: Wash the air filter cover regularly to prevent any
blockage in the cover. Do not boil. Make sure the cover is dry
before inserting the new air filter.
3. Insert a new air filter into the air filter cover.
CAUTION: Before inserting the new air filter, make sure the air
filter is clean and free of dust. Do not operate the device without
the air filter.
4. Put the air filter cover back on the compressor.
5. Do not maintain or service the device while it is in use.
TROUBLESHOOTING GUIDE
PROBLEM
CAUSE
No power on unit
The AC power cord is not
Turn the power switch off. Plug
when the power
plugged into an electrical
the power plug into an electrical
switch is on.
outlet.
outlet. Turn the device on.
No medication in the medi-
Add the correct amount of
cation tank. Too much or
prescribed medication to the
too little medication in the
medication tank.
medication tank.
The nebulizer kit is not
Make sure the nebulizer kit is
correctly assembled.
correctly assembled and the
inhalation accessory is correctly
attached.
No nebulization or
The nebulizer kit is tilted at
Hold the nebulizer kit correctly.
low nebulization
an incorrect angle.
Do not tilt the nebulizer kit so the
rate when the
angle of the kit is greater than
power is on.
45 degrees.
Make sure the air tube is correctly
The air tube is incorrectly
attached.
attached to the compressor and
the nebulizer kit.
The air tube is folded or
Make sure the air tube is not
damaged. The air tube is
folded, kinked or bent. Inspect the
blocked.
air tube for any damage. Replace
the air tube if damaged.
The device is
The compressor is
Do not cover the compressor with
very hot.
covered. The device has
any type of cover during use. Turn
been used for longer than
the device off. Wait 50 minutes
50 minutes.
before using the device again.
SPECIFICATIONS
Input
AC 120 V, 60 HZ
Extreme Pressure
>205 kpa
Power Consumption
≤200 VA
Medication Capacity
10 mL
Particle Size
≤5 µm & 60%
MMAD
2.0 µm
Average Nebulization Rate
≥0.2 mL/min
Noise Level
≤ 60 dB
Compressor Pressure Range
30-58 psi
Operation Pressure Range
7-10 psi
Liter Flow Range
6-10 lpm
Operation mode
50 minutes on, 50 minutes off
Operation Temperature Range
50˚F to 104˚F (10˚C to 40˚C)
Operation Humidity Range
30 to 75% RH
Storage Temperature Range
14˚F to 149˚F (-10˚C to 65˚C)
Storage Humidity Range
10 to 95% RH
Accessories
1 Infant Mask, 1 Pediatric Mask,
1 Reusable Nebulizer Cup,
1 Mouthpiece, 1 T-Piece,
1 7' (2.1 m) Tube, 5 Filters
Dimensions
7.1" × 6.3" × 9.1"
5
Equipment Classification with respect to protection
from electric shock: CLASS II
Degree of protection from electric shock: TYPE BF
Degree of protection against ingress of water is rated as IPX0.
Equipment not suitable for use in the presence of flammable
anesthetic mixture with air, oxygen or nitrous oxide.
Intermittent operation: 50 minutes ON, 50 minutes OFF.
LIMITED WARRANTY
SOLUTION
Your Dynarex Product is warrantied to be free of defects in
materials and workmanship for three (3) years for the original
purchaser. This item was built to exacting standards and carefully
inspected prior to shipment. This Limited Lifetime Warranty is an
expression of our confidence in the materials and workmanship
of our products and our assurance to the consumer of years of
dependable service.
The Warranty shall not apply under the following conditions:
Problems arising from normal wear
Problems arising from failure to adhere to the product
instructions
Problems arising from misuse, negligence, accident or
improper operation, maintenance or storage
Problems arising from modifications or unauthorized
repairs, parts or attachments
Products where the serial number has been removed
or defaced
Problems with non-durable components, such as rubber
accessories, casters, and grips, which are subject to normal
wear and need periodic replacement
Dynarex shall not be liable for any consequential or incidental
damages whatsoever. Dynarex shall repair or replace defective
products at its option. The foregoing warranty is exclusive and
in lieu of other express warranties, if any, including the implied
warranties of merchantability and fitness of a particular purpose.
The remedy for any violation of the implied warranty shall be
limited to repair or replacement of the defective product pursuant
to the terms contained herein.
If you have a question about your Dynarex device or this
warranty, please contact an authorized Dynarex dealer.
95%
14˚F
10%
-10˚C
Manufactured for:
Dynarex Corporation
10 Glenshaw Street
Orangeburg, NY 10962
USA • www.dynarex.com
Made in China
R210429