LED display reads
Pneumatic system leakage or cuff is too loose during inflation
"Er4"
LED display reads
Cuff pressure above 300mmHg
"Er5"
LED display reads
More than 160 seconds with cuff pressure above 15 mmHg
"Er6"
LED display reads
memory accessing error
"Er7"
LED display reads
Device parameter checking error
"Er8"
LED display reads
Pressure sensor parameter error
"ErA"
LED display reads
Bluetooth communicate error
"
Er"
The cuff position was not correct or it was not
properly tightened
LED display reads
an abnormal
Body posture was not correct during testing
result
Speaking, moving arm or body, being angry,
excited or nervous during test
Bluetooth
Bluetooth connection unsuccessful, monitor is
connection
abnormal, or strong electromagnetic interference
unstable
is present
No response when you
Incorrect operation or strong electromagnetic interference
press button
CARE AND MAINTENANCE
1. If this monitor is stored near freezing temperatures, allow it to return to room temperature before use.
2. If the monitor is not used for a long time, please be sure to fully charge it every month.
3. No monitor component needs to be maintained by the user. The circuit diagrams, component part lists, descriptions, calibration
instructions, or other information which will assist the user's appropriately qualified technical personnel to repair those parts of
the equipment which are designated for repair can be supplied by the iHealth technical department.
4. Clean the monitor with a dry, soft cloth or a moistened and well wrung soft cloth using water, diluted disinfectant alcohol, or
diluted detergent.
5. The monitor can maintain the safety and performance characteristics for a minimum of 10,000 measurements or three years of
usage, and the cuff integrity is maintained after 1,000 open close cycles.
6. The battery can maintain the performance characteristics for a minimum of 300 charge cycles.
7. It is recommended that if the cuff is used in a hospital or a clinic, it be disinfected twice a week. Wipe the inner side (the side
that contacts skin) of the cuff with a soft cloth lightly moistened with Ethyl alcohol (75-90%). Then air dry the cuff.
8. It is recommended that product performance be checked every 2 years or after each repair. Please contact the
iHealth Customer Service Center to do so.
Do not drop this monitor or subject it to strong impact.
Avoid high temperature and direct sunlight. Do not immerse the monitor in water as this will result in damage to the
monitor.
Measure again after five minutes. If the monitor is still abnormal, please
contact the local distributor or the factory.
Connect the mobile device correctly and try again, If the monitor is still
abnormal, please contact the local distributor or the factory.
Review the cuff application instructions and retest
Review body posture instructions and retest
Retest when calm; avoid speaking or movement during the test
Reset iOS/Android device. Reset monitor by pressing the START/STOP
button and holding for about 10 seconds. Make sure the monitor and
iOS/Android device are away from other electrical equipment. Please see
GENERAL SAFETY AND PRECAUTIONS
Press the START/STOP button and hold for about 10 seconds to reset the
device.