GaleMed MR-100 Plus Руководство по эксплуатации - Страница 8

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Респираторный продукт GaleMed MR-100 Plus. GaleMed MR-100 Plus 11 страниц.

Specifications 
Dead apace: 
Non‐rebreathing valve   
7ml
Adult mask 
150ml
Child mask 
9 5 m l
Infant mask 
2 8 m l
 
Pressure Relief 
4 0 ± 5 c m H
 
60±1 0 c m a H
Reservoir volume   
2 0 0 0 m l (adult model)
 
1 0 0 0 m l (c hi l d model)
 
5 0 0 m l (Infant model)
Bag volume/Stroke volume 
 
1 5 0 0 / 1 3 5 0 m l ( a d u l t m o d e l )
 
6 0 0 /
 
2 8 0 /
 
Maximum BPM 
4 5 b r e a t h s / m i n ( a d u l t m o d e l )
 
1 0 5 b r e a t h s / m i n ( c h i l d m o d e l )
 
9 8 b r e a t h s / m i n ( I n f a n t m o d e l )
 
Oxygen Concentration 
With reservoir 
9 9 %
Without reservoir 
4 5 % ( a d u l t a n d c h i l d m o d e l )
9 0 % ( i n f a n t m o d e l )
 
 
0 (infant & child)
2
0 (adult)
2
3 5 0 m l ( c h i l d m o d e l )
1 6 0 m l ( i n f a n t m o d e l )
Testing The Resuscitator 
The MR‐100™ Plus    Manual Resuscitator should be tested as follows: 
when first using the new Resuscitator 
after cleaning and disinfection 
after any new parts having been fitted 
monthly, if the Resuscitator is not frequently used. 
 
Equipment required: Test lung, 0‐100cmH
Child resuscitators only) , flow meter, regulated gas supply & gas supply 
tubing. 
 
Testing the silicone bag assembly 
1. Remove the non‐rebreathing valve and the oxygen reservoir (if fitted). 
2. Compress the silicone bag and occlude (block) the non‐rebreathing valve 
outlet. 
3. Release the bag. The bag should expand immediately and refill. If not, 
check the intake valve at the base of the silicone bag is correctly 
assembled or not. 
4. While keeping the non‐rebreathing valve outlet blocked, compress the     
bag again. The bag should not be compressed easily If this occurs, check 
that you are blocking the valve sufficiently, and that the intake valve at 
the base of the silicone bag is correctly assembled. 
 
Non‐rebreathing valve assembly 
1. Connect the non‐rebreathing valve to the silicone bag. Connect the test     
lung to the outlet on the non‐rebreathing valve and is correctly 
assembled. 
0 manometer (for Infant and     
2