Game ready 550550 Руководство пользователя - Страница 6
Просмотреть онлайн или скачать pdf Руководство пользователя для Блок управления Game ready 550550. Game ready 550550 13 страниц.
d e S C r i P t i o n
Control Unit
Heat Exchanger
Sleeve
Game ready System: The following items are included in your Game Ready System: a
Control Unit, an AC Adapter, a Power Cord, a Connector Hose, and a Welcome Packet,
containing the User's Manual and other information about the System. Each Wrap (comprised
of an inner Heat Exchanger and an outer Sleeve) is sold separately and is not included in the
System, however a Wrap must be attached to the System to begin treatment.
Pressure Backlight
Pressure
Power
Program
Play/
Pause
Subtract
Time
(Manual Mode)
ModeS oF oPerAtion
Manual Mode: The System defaults to this mode, and allows the user to adjust time and
pressure settings.
Program Mode: This mode allows the user to choose one of six programs that provide
treatment for a set time then sleep (no treatment) for a set time, continuously, at a specific
pressure setting.
8
o F
d e v i C e
Power Cord
AC Adapter
Welcome
Packet
Connector
Hose
Wrap
Volume
Pressure
Indicator
Average Wrap
Temperature
Average
Off
Low
Med
High
Wrap Temp.
55/13
50/10
45/7
Temperature
40/5
Knob
ºF/ºC
Sleep
Run
Coldest
Time
Time
Wrap Temp.
Coldest Wrap
C/F
(Celsius/Fahrenheit)
Temperature
Add
Time
drain Mode: This mode allows a user to connect a Hose with a special Hose Adapter
to the unit, enter Drain Mode and have the unit empty the water out of the ice box through
the Hose. Drain Mode can be accessed by pressing the program key and toggling through
all six programs. To empty water in the System while in Drain Mode, press the program
key until you reach Drain Mode, attach the Hose Adapter to the Connector Hose, place the
Hose Adapter over a sink, and press the play button. Drain Mode will run the Control Unit's
fluid pump for up to six minutes (long enough to fully empty the ice box). Drain Mode is
indicated by the following icons:
BUttonS
Power: Use this button to turn the Control Unit on and off.
Program: Use this button to select one of the available Programs or to return
to Manual Mode. See page 5 in this manual for more information on Programs.
Play/Pause: Use this button to start or pause a treatment or to stop.
Add time: Use this button to add time in Manual Mode (does not work in
Program Mode). You can add up to 90 minutes. Treatment must be paused
in order to add or subtract time.
Subtract time: Use this button to reduce time in Manual Mode (does not work in
Program Mode). You can subtract up to 90 minutes. Treatment must be paused in
order to add or subtract time.
Pressure Selection: Use this button to select one of four pressure settings:
No Pressure, Low Pressure (5-5 mmHg), Medium Pressure (5-50 mmHg), and
High Pressure (5-75 mmHg). Pressure selection is not available in Program Mode.
Treatment must be paused to change pressure settings.
volume: Use this button to select the option of sound or no sound. Alarms will still
sound.
C/F Button: Use this button to select either Celsius or Fahrenheit on
the temperature display.
Backlight: Use this button to turn the backlight on or off.
Adjusting temperature: The Control Unit cools the Wrap by pumping water from the
ice box into the Wrap; therefore, the temperature of the Wrap will never be colder than
the water in the ice box. If the knob is turned to three (3) snowflakes, the Control Unit
will provide the coldest possible therapy based on the temperature of the water in the ice
box. Turning the knob towards one (1) snowflake will provide less cold therapy. At one (1)
snowflake, the Control Unit is no longer providing cold therapy. At this temperature setting
the fluid pump will stop and the temperature will not be displayed.
note: The unit should always be started on full cold (3 snowflakes). The user can then
adjust temperature as desired.
Adjusting
temperature:
Temperature Knob
Turn knob counter-clockwise
for less cold therapy
More Cold
Less Cold
Full Cold
Turn knob clockwise
for more cold therapy
9