Operation
P MIN
P MAX
CAL ZP
16
Menu 6: Changing the minimum pressure indication
1.
Call up menu point "P MIN".
2.
Press OK.
"P MIN" starts to blink.
3.
Press
to set the currently present pressure as the minimum
value or
press
in order to set the value to Zero.
4.
Press OK to exit from this menu point.
Menu 7: Changing the maximum pressure indication
1.
Call up menu point "P MAX".
2.
Press OK.
3.
"P MAX" starts to blink.
4.
Press
to set the currently present pressure as the maximum
value or
press
to set the value to Zero.
5.
Press OK to exit from this menu point.
Menu 8: Calibrating the Zero point
Should you notice a deviation of the indicated value in relation to the
Zero point, the unit can then be recalibrated. A reference pressure is
necessary if the Zero point is different from ambient pressure.
1.
Call up menu point "CAL ZP".
2.
Press OK.
"CAL ZP" starts to blink.
3.
If the Offset is different from the ambient pressure, pressurize
the unit by applying a reference pressure. This pressure must
then correspond to the required starting pressure of the measur-
ing range.
4.
Press
to read the pressure.
5.
Press OK to memorize the displayed signal as the start of the
measuring range.
Should this configuration lead to a worsening of the original results,
for instance due to lack of a suitable reference pressure, the original
factory calibration as per original order can be re-instated via the
menu 10 "LD FAC".
DIM 20