Major tech MT957 Руководство по эксплуатации - Страница 6
Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Измерительные приборы Major tech MT957. Major tech MT957 17 страниц. Lcr meter
Symbols and Annunciators
HOLD
AUTO
MANUAL
SPECIFICATIONS
The instrument
complies with:
Insulation:
Overvoltage
category:
Display:
Polarity:
Overrange:
Low battery
indication:
Measurement
rate:
Auto power off: Meter automatically shuts down after
Operating
environment:
Storage
temperature:
Continuity
Low Battery
Diode
Data Hold
Auto Ranging
Manual Ranging
EN61010-1
Class 2, Double insulation
CATIII 600V
11000 counts LCD display with function indication.
Automatic, (-) negative polarity indication.
"OL" mark indication.
The "
" is displayed when the battery
voltage drops below the operating level.
4 times per second, nominal.
approx. 15 minutes of inactivity.
0 °C to 50 °C (32 °F to 122 °F) at < 70 % relative
humidity.
-20 °C to 60 °C (-4 °F to 140 °F) at < 80 %
relative humidity.