Tube placements Swing out rotor
This is not balanced!
E
ven though there are two samples in two buckets opposite to each other it is not balanced.
NEVER run swing out rotors with only two buckets
All four buckets must be used at all times
A
Wrong
Wrong
B
Always
use the same size tubes as in A
per bucket.
NEVER use different size tubes as in B
per bucket
Check they fit and do not hit the
centre as shown in C
C