2M Technology 2MTHWT-HMD Краткое руководство - Страница 9

Просмотреть онлайн или скачать pdf Краткое руководство для Металлоискатель 2M Technology 2MTHWT-HMD. 2M Technology 2MTHWT-HMD 12 страниц. Walkthrough metal detector
Также для 2M Technology 2MTHWT-HMD: Руководство пользователя (20 страниц), Руководство пользователя (8 страниц)

2M Technology 2MTHWT-HMD Краткое руководство
7. Measurement time: 0.5 seconds. After the measurement, please leave quickly. Otherwise the
measurement will be repeated.
8. During the temperature measurement, influenced by the measurement speed, the temperature might
suddenly fluctuate up and down. In this case, please measure the body temperature 3-5 times
continuously, and take the average value as the final measurement result.
9. The instrument automatically detects the temperature of the object in 20cm (7.87" in) in front of it. After
measuring the temperature, we have to move our hand or forehead away. If we keep close to the sensor
for long time, it will stop working as a protection.
Common trouble shooting
1. Metal detection does not count Processing method: press the power key to switch from the
temperature measurement interface to the metal detection and counting display interface.
2. The temperature cannot be measured when the machine is started Treatment: wait for 30 seconds after power
on, and then measure the temperature.
3. Low temperature measurement Processing method: check whether the value set in 1319 interface is 25, if not,
change to 25. And then use the hand-held electronic infrared Thermometer comparison test, if the measured
values are consistent, it means that the equipment is working normally. If there is a deviation greater than 0.5 ℃,
please refer to Debugging is described on the second page.
4. Temperature measurement is not sensitive Treatment: when measuring body temperature, stay at the
target measured by infrared temperature sensor for 1 second.
5. Metal detection sensitivity is too high Processing method: the setting item L008 in function setting password 1111
is changed from 1 to 50, excluding small metal parts.
6. Network connection failure Processing method: check whether the IP address of the computer is completely
consistent with the instructions. If not, please set the IP address according to the instructions, if it is
consistent, please refer to the video and operate again.
Body Temperature Measurement – Quick Guide V 6.1