AFS OLK51MM Руководство пользователя - Страница 6

Просмотреть онлайн или скачать pdf Руководство пользователя для Измерительные приборы AFS OLK51MM. AFS OLK51MM 15 страниц. Olk51 series complete fiber optic testing solutions

AFS E
UROPE
A
DVANCED FIBER SOLUTIONS

OM120A Operating instructions

Control Key's: (All key's are rubber key pads)
1. On\Off
2. Backlight
3. Wavelength Select ‫058( ג‬nm, 1300nm, 13l0nm, l490nm and l550nm)
4. dB\dBm\W
5. Zero reference (for all specified wavelengths)
Backlight Key allows the user the option to turn the display backlight on or off.
The Wavelength select key ‫ ג‬allows the user to select between the calibrated wavelengths of the OM120A.
All wavelengths are calibrated to ± 0.2 dB of the NIST standard.
The dB\dBm\W select key allows the user to select between absolute power, which is measured in dBm,
or relative power that is measured is dB. dBm\W is referenced to 1 milliwatt and both are a function of the
log scale.
The Zero reference key when in the dBm mode allows the users to toggle the resolution between 0.0ldB
and 0.ldB
The Zero reference key when in the dB mode allows the user to set the input launch power to a relative
dB zero reference value for easy calculation of loss and attenuation of a fiber Optic cable. For instance, if
a launch cable output power value is -25.6 dBm and a test cable is inserted between the launch cable and
the meter with a reading of -27.1 dBm, the user has to subtract -27.1 dBm from -25.6 dBm to get a result
of -1.5 dB. As simple as this test sounds it can sometimes get confusing to the user due to the fact that
they are not sure if their result should be a positive value or a negative value also miscalculation can be an
issue. The zero reference function eliminates these problems. The user simply sets their reference value
(which in this case is -25.6 dBm) to 0.0 dB then inserts the test cable to get a loss reading of -1.5 dB.
*EASY*.
Different references can be stored for all four wavelengths. The stored reference value will be maintained
until a new reference is stored.
The screen also displays the value that the user has referenced too at the bottom left corner of the screen.
This reading is the dBm (absolute) value of the users dB zero reference value.
* The OM120A also offers an auto power down feature. This function can be enabled or disable when the
unit is powered up by pressing the ‫ ג‬key. When enabled the unit will shut down after 15 minutes of no key
activity. The unit will signal the user that it is about to shut down by counting down the last minute in the
right top corner of the screen.
Tel:
00-31-342-470734
Fax:
00-31-342-422616
Video: 00-31-342-477888
:
Hunnenweg 7
Terms of payment and
3781 NM Voorthuizen
delivery published at Chamber
Postbus 111
of Commerce Harderwijk
3780 BC Voorthuizen
K.v.K. nr. 08077470
Rabobank
BTW/VAT-number
36.63.12.553
NL121102609B01
Postcheque BEF
Internet
000-1692940-96
www.afs-europe.com