Starting from the home page, press the F2 key to
enter the Service Menu.
Use the F2/F3 keys to select the "Calibration Menu"
option, then confirm with F4.
Access through the following password:
"F8; F7; F6; F8".
Select, using the F3/F2 keys, the "Detector
Calibration" option and confirm with the F4 key.
If required, press F3 to be able to enter the serial
numbers of the detectors so that they can
eventually be mentioned in the print report.
If necessary, also enter the serial number of the tool
and the name of the operator.
Continue by pressing F4.
Page 10/14
Cod. R0368 ‐ rev.1.0
Figure 17
Figure 18
Figure 19
Figure 20