Error Message
Snr
rEP
H
Lo
Err
Follow these directions to recall your last test result. From the main
screen (temperature display) press the Left Button one (1) time. Your
last test result will display on-screen.

SET TEMPERATURE DISPLAY FORMAT

Follow these steps to set the temperature display format to Fahrenheit
°F or Celsius °C. The default setting is Fahrenheit °F.
Step 1) From the main screen (temperature display) press the Right
Button for two (2) seconds until "USr" is displayed on-screen.
Step 2) Press the Power Button one (1) time.
Step 3) Press the Left Button to choose your desired format:
Cause and Course of Action
There has been an error with the sensor. This can occur
if the test subject does not wait at least 20 minutes after
smoking, eating or drinking before using the unit or if
the unit has not been used for an extended period of
time. Press the Power On/Off button once to turn off the
unit and repeat operation steps 1 – 5. If this message is
displayed four or more consecutive times, the unit may
need to be serviced.
The unit did not detect an even alcohol concentration in
the breath sample. This can occur if the user varies the
blowing force when performing a test or if the exhaust port
is obstructed. Repeat operation steps 1 – 5 to perform
another test.
The unit detected a very high alcohol concentration
(above 0.40% BAC).
Low Battery. Install two (2) new AA batteries.
The sensor has sustained damage; please contact us for
further support.

RECALL LAST RESULT

7