A.H. Systems 2383 Руководство по эксплуатации - Страница 9

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Антенна A.H. Systems 2383. A.H. Systems 2383 15 страниц. Active loop antenna

A.H. Systems 2383 Руководство по эксплуатации
A.H. Systems Model SAS-562B Active Loop Antenna
GENERAL USE INSTRUCTIONS
The SAS-562B loop antenna is an active loop designed to perform magnetic
field and shielding effectiveness testing. The loop antenna has a balanced
Faraday shield to reduce the E-field response for a pure magnetic field
measurement.
Each antenna is individually calibrated per IEEE 291-1991 and comes with a
calibration certificate references the NIST traceable test equipment. Initially
the antenna comes with a 1 year calibration interval, however depending on
the type of use and standards that the antenna is being used, this interval can
be adjusted.
The calibration tables shown provide a listing of the frequency of operation
and its antenna factor in dB/m. The field strength is the receiver voltage in
dBV plus the antenna factor (refer to the antenna factor calibration) plus any
cable loss. When making a measurement, mount the antenna on an
appropriate mast or tripod with the plane of the loop antenna facing the DUT.
 A.H. Systems inc. – Oct 2014
REV B
9