AB Sciex API 5000 LC/MS/MS Руководство по аппаратному обеспечению - Страница 9

Просмотреть онлайн или скачать pdf Руководство по аппаратному обеспечению для Лабораторное оборудование AB Sciex API 5000 LC/MS/MS. AB Sciex API 5000 LC/MS/MS 16 страниц.

AB Sciex API 5000 LC/MS/MS Руководство по аппаратному обеспечению
Hardware Guide
BACKING PUMP
CURTAIN GAS SUPPLY
MAX 60 PSIG
EXHAUST SUPPLY
GAS 1 / GAS 2
EXHAUST WASTE OUT
MIN 55 PSIG
MAX 105 PSIG
MAX 60 PSIG
Figure 1-6 Gas and vacuum panel

Control Panel

The control panel is located on the right side of the instrument. It encloses the
power distribution module.
The control panel contains the main power switch, four fuses, and two
Indicator/Reset buttons. The Vacuum Status button (green) and the Fault button
(red) indicate the vacuum status on the system controller board. When triggered
together, simultaneously pushed and released, the buttons reset the two
microprocessors on the system controller. This has an effect similar to restarting
the computer.
CAUTION! Resetting the instrument using these buttons interrupts any scans and
results in the loss of any unsaved files and data.
When the operational vacuum conditions are satisfied and the instrument is in
analysis mode, the Vacuum Status button is illuminated and the Fault button is
extinguished. The Fault button flashes when a vacuum fault is detected. In Pump-
Down mode, the buttons indicate the status of the Pump-Down sequence.
Document Number: D1000092206 B
9