TROUBLESHOOTING
The TrueView™ 1100 is running normally when no alarm message is present on the screen. If
an alarm appears, perform the action below to resolve the issue. Contact Technical Support if
you need additional help.
1.
Alarm Messages:
OUT OF RANGE: Once calibrated the TrueView™ 1100 will indicate an over range or
•
under range condition by setting the alarm. The error will remain active if -20% or 120%
of calibrated tension is exceeded.
Action Required:
new calibration will need to be established to increase the measurement range and
prevent future alarms.
WIRING ERROR: Will alert until the load cells are wired correctly.
•
Action Required:
terminal blocks, wiring shorts and ensure load cells are connected on the opposite end
of the cable. If the transducers need trouble shooting – contact tech support for
assistance.
LT Transducer: A WIRING ERROR indicated with an LT transducer could also
be caused by an overload condition.
Action Required:
rating. Reduce wrap angle to reduce effective net force exerted on load cell.
EXCITATION SHORT: Wiring or load cell failure.
•
Action Required:
need troubleshooting – contact tech support for assistance.
Action Required:
To clear this error the tension must be brought back into range. Or a
Check wiring and retry. Refer to page 3 to check for loose wires at the
Check that the tension range does not exceed the transducer load
Check for shorts in the transducer (load cell) wiring. If the transducers
Contact DFE for load cell replacement.