5. Insert electrode in buffer solution pH 4. After stabilization of the measured value, set the
display to pH 4 (No. 3).
6. Clean the probe with distilled water and put it into the buffer solution pH 7 again. Wait until the
pH value is stable. If pH 7 is not reached, repeat the calibration procedure several times.
The unit is ready for use after successful calibration.
To ensure a trouble-free use, the unit should be recalibrated and the probe be cleaned at least
every 4 weeks. Measuring with a pH-controller is only as exact as the calibration. The electrode
shall be emerged into water at most that the top part remains dry. Best is using the Aqua Medic
electrode holder.
The unit must be protected against moisture, splash or condensation water.
5. Troubles
-
The unit shows wrong values on the display:
Recalibrate the electrode.
-
The electrode cannot be recalibrated:
If the electrode cannot be recalibrated, it has to be renewed. pH electrodes have only a limited
life span, depending on kind of use between 1 and 3 years.
6. Technical Data
Measuring range:
Resolution:
Working temperature:
Humidity:
Power connection:
7. Warranty
Should any defect in materials or workmanship be found within 24 months of the date of purchase
(this is for the measuring unit, 6 months for the pH electrode) AB Aqua Medic undertakes to repair,
or at our option replace, the defective part free of charge – always provided the product has been
installed correctly, is used for the purpose that was intended by us, is used in accordance with the
operating instructions and is returned to us carriage paid. Proof of Purchase is required by
presentation of the original invoice or receipt indicating the dealer's name, the model number and
date of purchase. This warranty may not apply if any model or production number has been
altered, deleted or removed, unauthorised persons or organisations have executed repairs,
modifications or alterations, or damage is caused by accident, misuse or neglect. Please note that
the product is not defective under the terms of this Warranty where the product, or any of its
component parts, was not originally designed and / or manufactured for the market in which it is
used. These statements do not affect your statutory rights as a customer.
AB Aqua Medic GmbH - Gewerbepark 24 - 49143 Bissendorf/Germany
Current manuals are available for download at www.aqua-medic.de.
6
pH 0 - 14
0.01 pH
0 – 50 °C
below 80%
input: 100 – 240 V, 50/60 Hz AC, output: 12 V DC
- Technical changes reserved - 03/2018