Allied Radio Knight-Kit 83 Y 140 Руководство по сборке - Страница 6
Просмотреть онлайн или скачать pdf Руководство по сборке для Оборудование для звукозаписи Allied Radio Knight-Kit 83 Y 140. Allied Radio Knight-Kit 83 Y 140 10 страниц.
With your KNIGHT
VOM you can perform a very
great variety of measurements.
The different measure-
ments are selected (a) by different panel jacks into
which the test leads are inserted and (b) by the posi-
tion of the Function Switch.
A number of scales on
the meter allow proper readings for any range of any
function provided.
The Function Switch has 18 posi-
tions and 2 decks.
This makes it possible to connect
the different components of the instrument so that d-c
and a-c voltages, direct and alternating currents, and
resistances can be measured.
The meter has a highly
sensitive 50-ua moving coil
permanent magnet move-
ment.
A calibrating resistor
(R3) assures the highest
degree of accuracy of the dial indications.
The meter
serves as the indicator for all measurements.
It has
separate scales for a-c voltage, d-c voltage, decibels,
xesistance and current.
Preliminary
Adjustments
Before
any measurements
are made,
be sure
the
instrument is placed squarely on a bench
where the
measurement is to be performed.
Always check that
the meter pointer is lined up with the zeros on the
left hand side.
If the pointer does not fall in line
with
these marks,
turn the bakelite
screw directly
above the word "'OFF"' either left or right until pointer
is positioned correctly.
Another reminder before using the meter for a par-
ticular type of measurement
is to make certain that
the operator understands fully how to handle this in-
strument,
as well as its capabilities.
With this in
mind, the following paragraphs are written, to give an
outline on how to use the KNIGHT VOM to your best
advantage.
How to Measure
D-C Voltoges
Coution:
Wherever possible, disconnect power
from
the equipment under test before connecting
the meter.
In cases where it is impossible to do
this, connect the leads separately and use only
one_hand at any one time.
Grasp the test prod
well back from the metal tip and make momentary
contact with the circuit under test by letting the
metal tip of the prod touch the point at which the
potential
exists.
Never
touch
any part of the
equipment under test with the other hand.
Plug the black test lead into the ""COM'' jack and
the red test lead into the ""V. ,
.A"' jack. First touch
the probe end of the black (negative) test lead to the
negative
side of the circuit to be checked.
Then,
leaving one hand free, touch the red (positive) test
probe
to the positive
terminal
of the circuit to be
tested.
It is a good practice
always
to start with the
highest
range
available,
as
a protection
for your
meter.
Then, after first indication, the switch should
be reset to the position in which more accurate read-
ings can be obtained.
If possible, make final readings
on a range on which the indication is in the right hand
half of the scale, because meter accuracy is based on
full scale value.
The
value of the measured
voltage can now be
read
from
the
appropriate
meter
scale,
taking into
account
the range
setting of the Function
Switch.
Read
on
tne meter
scale
whose
extreme
right-hand
Rn
marker is labeled with the full-scale value indicated
by the Function Switch or some multiple or submultiple
of it.
For d-c voltage readings, use only the black
scales marked "*DC''.
For the 0-2.5 v d-c range use the 250 v d-c scale.
Divide meter scale reading by 10.
Each minor scale
division equals 0.05 v d-c.
For the 0-10 v d-c range use the 10 v d-c scale.
Read scale directly. Each minor scale division equals
-2v d-c.
For the 0-50 v d-c range use the 50 v d-c scale.
Read scale directly. Each minor scale division equals
Iv de.
For the 0-250 v d-c range use the 250 v d-c scale.
Read scale directly.
Each minor division equals 5
v d-c.
For the 0-1000 v d-c range use
10 v d-c scale.
Multiply
scale reading by 100.
Each
minor scale
division equals 20 v d-c.
For the 0-5000 v d-c range use 50 v d-c scale.
Multiply reading by 100.
Each minor scale division
equals 100 v d-c.
If d-c voltages in excess of 1000 v but less than
5000 v are to be measured,
the red test lead has to
be plugged into the '5000 V DC' jack.
If working
with high potentials of this order, make doubly sure
that the Caution
paragraph at the beginning of this
section is strictly observed.
How to Measure A-C Voltages
NOTE:
Follow
same
precautions
as under
"Caution"?
in above
section
''How to Measure
D-C Voltages.
''
Plug the black test lead into the "'COM'' jack and
the red test lead into the "'V, .A''
jack.
Clip the
probe
of the black
test lead
to one
terminal
of the
potential to be measured and the probe end of the red
test lead should be touched
to the other terminal of
the potential.
Always
start
with
the highest
"'AC
V'' range
available, as a protection for your meter.
Then after
obtaining
the
first
indication,
reset
the
Function
Switch
to the position in which
a convenient meter
reading, preferably in the upper half of the scale, is
obtained.
Tum
the power on and observe the meter deflec-
tion.
The value of the measured voltage can now be
read
from
the
appropriate
meter
scale
taking into
account the range setting as indicated by the Function
Switch.
For the 0-2.5 v ac range use 2.5 v a-c scale.
Read scale directly.
Each minor scale division equals
0.05 v ac.
All other a-c voltage scales are chosen and read
in the same
way as the d-c scales discussed in the
previous
section,
except that the red "'AC'' scales
are used.
If a-c voltages in excess of 1000 v, but not higher
than 5000 v a-c are to be measured,
the red test lead
has to be plugged into the '5000 V AC" jack.
If
working with high potentials of this order, be doubly
sure that the Caution at beginning of previous section
is strictly observed.
19