Controls should be run when reagent packs are changed, after a Soak Cleaning has been
performed, and to check the calibration of the analyzer when results are in question.
Remove the control from the refrigerator and allow to warm to room temperature for
15 minutes before use. Mix control tubes thoroughly immediately prior to use. Roll 10X
between hands before gently inverting 10-15X.
Controls should be discarded if expired or if they have been open for more than 14 days.
1
3
Running Controls
1. From the Home Screen, touch the
Maintenence icon.
2. Touch the Quality Control.
3. Select the appropriate control level to match
the vial being used.
4. Touch Measure.
5. Touch Run QC.
6. If the lot number displayed matches the
control value sheet, check that the target
and gap values match the screen and
touch Accept.
If the lot number does NOT match,
touch New.