Abaxis VetScan UA10 Руководство - Страница 2

Просмотреть онлайн или скачать pdf Руководство для Измерительные приборы Abaxis VetScan UA10. Abaxis VetScan UA10 2 страницы. Urine test strips

b. A transfer pipette or syringe may alternatively be used to drop well-
mixed urine onto each pad, covering each pad completely – do not
leave urine on strip for more than 2 seconds (proceed to step 5
immediately). The latter method is preferable if cultures need to be
performed. If the sample has been refrigerated, test the urine only
after it has warmed to room temperature. (59-77 °F, 15-25 °C)
8. If using a dip method, remove the strip from the tube, dragging the edge
of the strip against the rim of the urine container to remove excess urine.
9. Blot the strip on the long edge to clean, absorbent paper to remove
excess urine to prevent cross-contamination from adjacent reagent pads.
Avoid contacting the surface of the pads directly with the absorbent paper
to prevent contamination from the paper.
10. Place strip on UA strip tray with the end of the strip at the edge of
the analyzer opening. The tray will slide into analyzer automatically
approximately 30 seconds after the Test button was pressed. Results
will appear onscreen in another 30 seconds, for a total test time of
approximately one minute.
PRECAUTIONS AND WARNINGS
1. The test is intended for veterinary use only.
2. The strip is NOT intended to be read visually, but only with a VetScan
UA analyzer.
3. Collect a fresh urine specimen in a clean, dry container or syringe. Do
not expose urine specimens to sunlight as this induces oxidation of
bilirubin and urobilinogen and hence leads to artificially lower results for
these two parameters.
4. If the sample will not be tested immediately, it may be stored at 59-77 °F,
(15-25 °C) for up to one hour before testing. If it will not be tested within
1 hour of collection, the sample may be refrigerated up to 4 hours.
When testing refrigerated samples, be sure to warm the sample to room
temperature (at least 15 minutes on counter or warm in palms of your
hand) (59-77 °F, 15-25 °C) prior to testing. Do not test cold urine. Be
sure not to freeze the sample.
5. Only use strips that are at room temperature (59-77 °F, 15-25 °C). If they
have been refrigerated, allow them to warm to room temperature before
use. Only perform tests at room temperature 59-77 °F, 15-25 °C.
6. Use gloves when handling the strips to avoid touching the reaction pads
on the reagent strips to prevent contamination.
7. Do not remove the desiccant pouch from the tube of strips. Replace
cap immediately and tightly after removing a reagent strip. Exposure
of unused strips to humid air (more than 5 minutes) can easily lead to
inaccurate test results.
8. Do not use reagent strips after the expiration date, and do not use
deteriorated, discolored or blackened test strips.
9. False-positive readings for blood and glucose can result from residues of
strongly oxidizing disinfectants in the specimen collection vessel. Do not
add preservatives to the urine. Avoid contamination by volatile chemicals.
10. The used test strip cannot be reused, but should be disposed of as
biohazardous waste.
11. When cleaning the strip tray, use alcohol wipes or a tissue moistened
with a mild soap and water. Dry tray thoroughly before replacing it into
the UA analyzer.
LINEAR RANGE Conventional Units (SI Units)
Analyte
Range
15-500 Cell/μL
Leukocytes
(15-500 Cell/uL)
0.5-8.0 mmol/L
Ketone
(2.9-46.5 mg/dL)
Nitrite
+
2.0-8.0 mg/dL
Urobilinogen
(33-131 μmol/L)
0.5-6.0 mg/dL
Bilirubin
(8.6-100 μmol/L)
PLEASE NOTE
In principle, diagnosis or therapy should not be based on one test result
alone but should be established in the context of all other medical findings.
Knowledge of the effects of drugs or their metabolites upon the individual tests
is not yet complete. In doubtful cases, it is therefore advisable to repeat the
test after discontinuing a particular drug. Large amounts of ascorbic acid in the
urine can produce artificially low to false-negative results for glucose, blood,
nitrite and bilirubin.
STORAGE AND STABILITY
Strips in the tube may be stored from 36-86 °F (2-30 °C). Abaxis recommends
to store at room temperature 59-77 °F (15-25 °C) as strips should be gently
warmed or cooled to room temperature, as appropriate, prior to opening the
tube to retrieve a test strip. Store only in original tube with desiccant pack,
avoiding humidity, direct sunlight and heat. NOTE: Be sure to tightly recap
the tube immediately after removing strips for use.
EXPIRATION
Unopened tubes of strips are usable until the expiration date. NOTE: Unused
strips that remain in the original capped container with the desiccant are stable
for 3 months after first opening of the tube. After the tube has been opened,
the strips must be used within 90 days (3 months).
PART NO.
1500-0013-50 VetScan UA10 Urine Strips, 50 strips/vial.
1500-0013-25 VetScan UA10 Urine Strips, 25 strips/vial
EXPLANATIONS FOR SYMBOLS ON THE LABEL
2
Temperature limit
Do not re-use
Consult instructions for use
Manufacturer
Date of Manufacture
Keep Dry
Good for 3 months after first opened
For veterinary use only
FOR VETERINARY USE ONLY
California Prop. 65 WARNING:
State of California to cause birth defects or other reproductive harm. Bisphenol
A CAS-No. 80-05-7. Revision Date 2017-09-27
Abaxis, Inc. 3240 Whipple Road, Union City, CA 94587
USA | +1 800 822 2947 | www.abaxis.com
ABAXIS Europe GmbH, Bunsenstr. 9-11, 64347 Griesheim,
Germany | +49 6155 780 210
Manufactured for Abaxis in China.
Abaxis VetScan products are intended for veterinary use only.
Abaxis and VetScan are registered trademarks of Abaxis, Inc. © Abaxis 2017.
Analyte
Range
50-1000 mg/dL
Glucose
(2.8-55 mmol/L)
15-300 mg/dL
Protein
(0.15-3.0 g/L)
Spacific Gravity
1.000-1.060 (same)
pH
5.0-9.0 (same)
10-200 Cell/μL
Blood
(10-200 Cell/uL)
Number of items included
X
Lot number
Part number
Keep away from Sunlight
Use-by-date
This product contains a chemical known to the
1500-7102 Rev. A