3B SCIENTIFIC PHYSICS 500 1013146 Руководство по эксплуатации - Страница 3

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Микроскоп 3B SCIENTIFIC PHYSICS 500 1013146. 3B SCIENTIFIC PHYSICS 500 1013146 4 страницы. Binocular microscope led with polarisation equipment

can be tested by rotating the polariser or the
specimen itself.
A birefringent object, when rotated continuously,
shows up brightly after each 90° rotation and is
dark between these positions. In contrast, ob-
jects that are isotropic and not birefringent re-
main dark in all positions.
4.3 Changing the fuse
Turn off the power switch and unplug the
mains plug.
Unscrew the fuse holder on the back of the
stand base with a screwdriver.
Replace the fuse and reinsert the holder in
its socket.
5. Storage, cleaning and disposal
Keep the microscope in a clean, dry and
dust free place.
When not in use always cover the micro-
scope with the dust cover.
Do not expose it to temperatures below 0°C
and above 40°C and a max. relative humid-
ity of over 85%.
Always unplug the mains plug before clean-
ing or maintenance.
Do not clean the unit with volatile solvents or
abrasive cleaners.
Do not disassemble objective or eyepieces
to attempt to clean them.
Use a soft linen cloth and some ethanol to
clean the microscope.
Use a soft lens tissue to clean the optics.
The packaging should be disposed of at
local recycling points.
Should you need to
dispose of the equip-
ment itself, never throw
it away in normal do-
mestic
waste.
regulations for the dis-
posal
of
electrical
equipment will apply.
3B Scientific GmbH • Rudorffweg 8 • 21031 Hamburg • Germany • www.3bscientific.com
Local
Subject to technical amendments
© Copyright 2013 3B Scientific GmbH