Aquatec Beluga Инструкции по сборке - Страница 3

Просмотреть онлайн или скачать pdf Инструкции по сборке для Инвалидное кресло Aquatec Beluga. Aquatec Beluga 4 страницы. Change of actuator
Также для Aquatec Beluga: Инструкции по сборке (4 страниц), Инструкции по сборке (2 страниц), Инструкция по сборке (3 страниц)

Assembly Instructions
Change of actuator Aquatec
5.
Remove the motor cable with gasket from the guiding.
6.
Push the actuator down again.
7.
Break the hooks of the motor bolt with the screw driver and remove the motor bolt.
8.
Now the actuator can be removed.
E.
ASSEMBLY
1.
Take the motor out of the packing.
2.
Replace the motor cable with gasket in the guiding.
3.
Carefully replace the gasket of the lid.
4.
Put the motor cable plugs on the connectors of the circuit board.
5.
Close the electronic casing with the lid. Make sure that the gasket is in a correct
position.
6.
Screw the lid on but please do not strip the screws!
7.
Put the new actuator in its position on the backrest. Now put in the new motor bolt
until it is engaged.
Status: 01.12.06 / Marketing
1470111 Change actuator Beluga
®
Beluga
Page 3 of 4
V/Allgemein/Montage/Montageanleitungen/Englisch
Beluga/Change actuator Beluga