SENSOR, ALARM AND OUTPUT CABLE CONNECTORS:-
~J
l]
Il
LJ
-;
~
~
LJ
To open the spring clamp terminals, EITHER:-
a)
Inseli a flat bladed screw-driver into the rectangular slot, above the cable entry
and twist it through 90°. This forces the cage open and the cable can be inserted
through the round hole.
Twist the screw-driver back through 90° and extract it from the rectangular slot,
to clamp the cable.
OR
b)
Only if the cable connector has been unplugged from the instrument, lay the
connector on a flat surface and push down on the upper section of the spring
clamp which is visible through the slot in the top of the connector. This forces the
cage open and the cable can be inserted through the round hole.
Remove the downward pressure on the spring clamp to clamp the cable.
SENSOR CABLE
Connect the sensor cable to the lower left 5 pin connector, ensuring that the correct
colour coding is observed and that the cage is securely clamped onto the bootlace ferrules
on the cable provided.
Root the sensor cable to the intended site of the sensor.
NOTE
Do not install the sensor at this time. Wait until the commissioning stage,
as described on page 9.
ALARM CABLE
Make the appropriate connections, noting the normally open and normally closed relay
contact positions.
ANALOGUE OUTPUT CABLE
Make the appropriate cOlmections, ensuring that the correct polarity and the maximum
load specification is strictly observed.
As can be seen from the label, the output can be used in either ACTIVE or PASSIVE
mode and it is impOliant that the correct option is used, to ensure normal operation of the
output system.